Relative Humidity
(continued)
Thermo Scientific
If a reliable RH measuring device is not available, the display can be
calibrated to a typical level;
1. Follow the RH stabilization periods outlined above.
2. With a full humidity pan and stable temperature, the relative humidity
in the chamber will be 95%.
3. Using Step 3-5 of the relative humidity sensor adjustment, adjust the
display to 95%.
This calibration method should be accurate to within 5%.
Section 2
Calibration
Direct Heat CO2 Incubator
2-5