●Device access lock parameters (index 12)
The contents are as follows.
Data type: 16 bit Record
Value (dec)
0x0000(0)
Key lock release, DS unlock (Initial value)
0x0002(2)
Key lock release, DS lock
0x0008(8)
Key lock, DS unlock
0x000A(10)
Key lock, DS lock
[Key lock]
This function prevents the user from physically changing the setting of the pressure switch (button
operation is not accepted).
Even when key lock function is activated, settings can be changed by IO-Link communication.
Restoration by data storage (overwriting parameter data) can be performed.
[Lock data storage (DS lock)]
Locking "Data storage" will invalidate the data storage function of the pressure switch.
In this case, access will be denied for backup and restoration of data storage.
●Device state parameters (index 36)
Readable device states are as follows.
Data type: 8 bit UInteger
Value (dec)
0x00(0)
Normal operation
0x01(1)
Maintenance inspection required
0x02(2)
Outside specification range
0x03(3)
Function check
0x04(4)
Failure
●Device detail status parameters (index 37)
Detailed event contents of readable device status are as follows.
Array
1
Internal failure of digital pressure switch
2
Internal failure of digital pressure switch
3
Internal failure of digital pressure switch
4
Internal failure of digital pressure switch
5
Internal failure of digital pressure switch
6
Internal failure of digital pressure switch
7
Internal failure of digital pressure switch
8
9
Device temperature upper limit exceeded
10
Measured pressure range upper limit exceeded
11
Fell below measured pressure range lower limit
12
13
14
Data storage upload request
Contents
State definition
Event content
-
-
-
-68-
Description
-
Not available
Device temperature upper limit exceeded
Measured pressure range upper limit exceeded
Falls below measured pressure range lower limit
Not available
Internal failure of digital pressure switch
Event classification
Definition
Value
0xF4
Error
0xF4
Error
0xF4
Error
0xF4
Error
0xF4
Error
0xF4
Error
0xF4
Error
0x00
-
0xE4
warning
0xE4
warning
0xE4
warning
0x00
-
0x00
-
0x54
notification
Event code
0x8D01
0x8D02
0x8D03
0x8D04
0x8D05
0x8D06
0x8D07
0x0000
0x4210
0x8C10
0x8C30
0x0000
0x0000
0xFF91