ENCODER PROTOCOL INTERFACE

The encoder protocol interface requires firmware version 1 .10 or later . Reference Badger Meter P/N 67354-003 to obtain a
firmware upgrade kit .
Enabling the meter as an encoder requires three settings, all within the advanced menu, to be configured .
Totalizer Resolution – Selects the resolution of the display totalizer .
Protocol Type – Selects the type of information to be transmitted to the encoder .
Dial Type – Enables encoder and selects the number of significant totalizer digits to transmit .
Changing the protocol type automatically configures the necessary digital inputs/outputs . Manually changing the digital
inputs/outputs within the Input/Outputs menu is not allowed . Below is a wiring diagram for connecting an encoder to
the meter .
N
OTE:
When connected to an ORION® Cellular LTE endpoint, add a 15K resistor (Badger Meter P/N 69224-001) to the meter
terminal block between terminal 1 (green wire) and terminal 8 (red wire) as shown to correct any potential meter
reading issues . The resistor is indicated by an arrow in the photo and in the drawing .
March 2022
Figure 38: Encoder interface
MAG-UM-01272-EN-21
Encoder Protocol Interface
(Badger Meter P/N 69224-001)