2-Point Adjustment - Optex TS-1000-D Instruction Manual

Ss checker
Table of Contents

Advertisement

During adjustment a progress bar is
displayed.
When completed, the display returns
to the measurement value display.

2-point adjustment

Display current values of the 2 points as any adjustment values,
respectively.
Performing 2-point adjustment changes the offset and span based on the
adjusted values.
For 4-20 mA output ("6 4-20 mA Setting" P.15) and alarm output ("Alarm
output setting" P.17), the adjusted values are applied.
Display value
after 2-point
adjustment
If offset, span, or 2-point adjustment has already been performed,
Caution
be sure to initialize the adjustment before performing the 2-point
adjustment.
Measurement
item [Display]
Suspended
solids [SS]
Transparency
[Trans.]
Press
1
The measurement item selection display is displayed.
[Point 1]
Adjusted
value
Current display
value
Measurement item setting range
(display value/adjusted value)
[Point 1]: 0.0 to (2nd point value - 20)
[Point 2]: (1st point value + 20) to 1000 1000
[Point 1]: 0 to (2nd point value - 20)
[Point 2]t: (1st point value + 20) to 100 100
.
[Point 2]
Adjusted value
Current display
value
Measurement
value
Factory default
setting
0.0
0
25