Upper/lower limit of signal output range

For the item by 4-20 mA signal output, set the measurement value to output
at lower limit (Lo, 4 mA) and upper limit (Hi, 20 mA).
Signal output
For operation, refer to the "Universal Transmitter SC-U1 Operation
Manual", "5.3 4-20 mA Setting", " ◆ Signal output range lower limit"
and " ◆ Signal output range upper limit".
Output item
[Display]
Suspended
solids [SS]
Transparency
[Trans.]
The setting unit is based on the display resolution of the item set for the
display item. (For display resolution, refer to "14 Specifications" P.34)
16
20 mA
4 mA
Measurement
value at 4 mA
Factory default
setting
Lower limit: 0.0
Upper limit: 1000
Lower limit: 0
Upper limit: 100
Measurement
value at 20 mA
Signal output range
Lower limit to Upper
limit
0.0 to 1000
0 to 100
Minimum setting
range
Upper limit - Lower
limit ≥
20.0
20