Voltage Calibration - Gain - Pacific Power Source AFX Series Operation Manual

Hide thumbs Also See for AFX Series:
Table of Contents

Advertisement

11.4.3 Voltage Calibration - Gain

Figure 11-1: Voltage Calibration Equipment Setup – 1 or 3 Phase Mode – Phase A
The procedure to calibrate voltage gain is:
1. Select the Calibration entry from the "System" menu
2. Press the "Calibrate" soft key to enter calibration mode
3. Do not connect any load to the output
4. Program AC voltage to full scale 300.0 Vrms L-N for AC mode calibration or 425Vdc for
DC mode calibration on all phases. Use the PROGRAM screen for this.
5. For AC mode calibration, set the Frequency to the most commonly used value
depending on typical applications
6. NOTE: If the user calibration is targeted for specific operating conditions, for example
Vac = 115V L-N and Freq = 400 Hz, use these settings instead of those suggested in steps
4 and 5
Entire Contents Copyright
2016 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
DMM
AFX SERIES® OPERATION MANUAL
SECTION 11: CALIBRATION
AC Line
Page 264 of 274
Input