11.4

Calibration Procedures

Calibration of output and measurements is performed by a single procedure so there is no need
to calibrate each separately. Calibration involves voltage and current full scale and offset on
each phase. For three phase mode, this means there are twelve calibration points, for single
phase mode, there are four.
The calibration can be performed manually using the Calibration menu (refer to Section 6.8.9 for
the location of the Calibration menu) or through the build-in web server using the web browser
interface. Refer to Section 10.10.7, "System Calibration" for access to the Web based calibration
screens.

11.4.1 Voltage Calibration - Offset

The procedure to calibrate voltage offset is:
1. Select the Calibration entry from the "System" menu
2. Press the "Calibrate" soft key to enter calibration mode
Entire Contents Copyright
2016 by Pacific Power Source, Inc. (PPS) • All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
DMM (optional)

Table 11-2: Setup for Voltage Offset Calibration

AFX SERIES® OPERATION MANUAL
SECTION 11: CALIBRATION
A B C N N N
Short A, B & C to N for
Voltage Offset
Calibration
AC Line
Input
Page 262 of 274