S p e c i f i c a t i o n s

SPECIFICATIONS

Weight ................................................................................14.7 ounces
16.3 ounces with battery
Dimensions ....4.75"L x 3.4"W x 2.8"H (12.06 cm L x 8.64 cm W x 7.11 cm H)
Power Supply ..........................(2) Standard "AA" alkaline or NiMH batteries.
Power Requirements............................Batteries other than alkaline or NiMH
are not recommended due to the capacity needed
for operation and battery life of the unit.
Operating Temperature Range..............................5° to 40°C (41° to 104°F)
Operating Pressure Range .........................500 to 2250 PSIG (34 to 155 bar)
tank pressure
Operating Atmospheric Conditions................................500 to 1020 millibar
Operating Humidity Range ...........................0 to 95% R.H., non-condensing
Storage and Transportation Temperature Range.........................-20° to 60°C
(-4° to 140°F)
Storage and Transportation Humidity Range ........................Up to 95% R.H.,
non-condensing
Degree of Protection Against Ingress of Liquids ..................................NONE
Degree of Protection Against Electric Shock...................TYPE BF applied part
Modes of Operation.....................................................Continuous / Pulsed
Approval Body And Safety Standard ......................IEC 601-1; CAN/CSA-C22.2
No 601.1-M90
US Patents .............................................................4,519,387; 5,755,224;
4,457,303