Devilbiss Ifill Personal Oxygen Station; Ifill Oxygen Cylinder/Regulator; Typical Questions And Answers; Cleaning/Maintenance Devilbiss Ifill Personal Oxygen Station Filter - DeVilbiss iFill 535D Instruction Manual

Hide thumbs Also See for iFill 535D:
Table of Contents

Advertisement

Available languages

Available languages

iFill Oxygen Cylinder/Regulator

PROBLEM
Hissing sound from iFill oxygen cylinder
iFill oxygen cylinder is on but oxygen is not being
delivered.

TYPICAL QuESTIONS ANd ANSWERS

Q. Why does my iFill make a "pop" sound when I remove the cylinder?
A.
The "pop" sound is pressure being released from the system to make the next cylinder easier to install.
Q. Why is the coupler only to be cleaned with a clean, dry, lint­free cloth and only when necessary?
A. C leaning solutions may be flammable and/or may contaminate the cylinder.
Q. Why can't I use my old cylinders with the iFill?
A.
The iFill has a proprietary connection as required by the Food and Drug Administration (FDA) and, therefore, other cylinders are not compatible.
CLEANING/MAINTENANCE
devilbiss iFill Personal Oxygen Station Filter
CAUTION – Do not operate the personal oxygen station without the filters installed.
NOTE– The DeVilbiss iFill Personal Oxygen Station has two (2) filters, one on each side of the cabinet (FIGURE A).
NOTE– Perform this procedure at least ONCE A WEEK.
1.
Unplug the personal oxygen station from the wall outlet.
2. Remove the filters.
3. Clean the filters with a vacuum cleaner or wash in warm soapy water and rinse thoroughly.
4. Dry the filters thoroughly before reinstallation.

Exterior Cabinet

WARNING
Electric Shock Hazard: Unplug the DeVilbiss iFill Personal Oxygen Station when cleaning. Do not remove the oxygen station cabinet. The cabinet should only be
removed by a qualified DeVilbiss technician. Do not apply liquid directly to the cabinet nor utilize any petroleum- b ased solvents or cleaning agents.
1.
Clean the exterior cabinet by using a damp cloth or sponge with a mild house hold cleaner and wipe it dry.
2.
Only if necessary, wipe the coupler with a clean, dry, lint-free cloth.
SPECIFICATI ONS

devilbiss iFill Personal Oxygen Station

Operating Temperature:.......................................................................................................................................................................................................... 41º to 95ºF (5º to 35ºC)
Operating Humidity Range: .......................................................................................................................................................................................15 to 95% R.H. non-condensing
Operating Altitude: ................................................................................................................................................................................................ 0 to 6,562 Feet (0 to 2,000 Meters)
Storage Temperature Range: ..................................................................................................................................................................................-40ºF to +140ºF (-40ºC to +60ºC)
Storage Humidity Range: .................................................................................................................................................................................................. 10 to 95% non-condensing
Electrical Rating: .................................................................................................................................................................................................................... 115V~, 60Hz, 4.1 Amps
Operating Voltage Range: ......................................................................................................................................................................................................... 187 – 264V~ 50/60Hz
Power Range: .................................................................................................................................................................................................................................400 Watts Average
Oxygen Purity: ....................................................................................................................................................................................................................................... 93% O2 ± 3%
Dimensions: (including casters) w/o cylinder ..................................................................................................................12.25"W x 28.5"H x 22.5"D (31.1 cm x 72.4 cm x 57.4 cm)
Weight: .................................................................................................................................................................................................................................................. 66 lbs. (30 kg)
Shipping Weight: ................................................................................................................................................................................................................................ 76 lbs. (34.5 kg)
This unit is classified as nonprotected per EN60601-1 .....................................................................................................................................................(Ordinary Equipment) IPX0
Approvals UL60601-1 & CAN/CSA C22.2 No. 601.1-M90 .................................................................................................................................................................................... CSA
®
C
US
156855

iFill Oxygen Cylinder Typical Fill Times

NOTE– All filling times are approximate and may vary depending on altitude and environmental conditions.
Typical cylinder fill time from empty to 2,000 ± 200 psig (138 ± 13.8 bar) are as follows:
M4 (0.7L) .................60 Minutes
M6 (1.0L) .................75 Minutes
ML6 (1.2L) ...............90 Minutes
NOTE- Degradation of performance may occur if unit is operated outside of specified operating parameters.
A-535D
1. Turn rotary selector OFF.
2. If hissing sound continues, call your DeVilbiss provider.
1. R efill cylinder.
2. Check that cannula tubing is straight and not pinched. If damaged, replace.
3. Call your DeVilbiss provider for service.
C (1.8L) .............130 Minutes
D (2.9L) .............215 Minutes
E (4.7L) .............350 Minutes
SOLUTION
EN - 7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents