English

SPECIFICATIONS

Catalog Number
Delivery Rate (Lower delivery rates
available for low flow applications)
Maximum Recommended Flow
Outlet Pressure
Electrical Rating
Operating Voltage Range
Oxygen Percentage
Operating Altitude
(tested at 21°C only)
0-1500 M (0-4921 ft)
1500-3000 M (4921-9842 ft)
3000-4000 M (9842-13123 ft)
Operating Environment Range*
10°C to 35°C, humidity range of 30% to 75%
Power Consumption
Weight
Sound Level (ISO 8359:1996 from front)
Dimensions
Pressure Relief Valve
Operating System
For units equipped with an OSD,
the visible "low oxygen" indicator
will activate at the following level
Storage Conditions
Equipment Class and Type
Approval Body
and
Safety Standard
EMC Compliance To:
*Note: The OSD performance at 10°C to 35°C, 95% R.H. through voltage range on the 515DS verified at 670m.
**Note: This corresponds with a sound level of 50 dbA overall average as measured according to the guidelines of the now defunct ANSI Z79.13-1981 standard.
Specifications subject to change without notice.
DEVILBISS 4-LITER SERIES
515UK
1 to 4 LPM
4 LPM
230 V, 50 Hz, 1.4 Amp
195-253 V~, 50 Hz
1-4 LPM = 93%±3%
Across the voltage range:
No degradation of performance
Not recommended/Not tested
Not recommended/ Not tested
No degradation in performance across
the operating voltage range
(tested at sea level).
295 Watts Average
53 lbs. (24.5 Kilograms)
50.5 dbA overall average
Units with serial numbers less than H20000: 85%±2% (The audible alarm will alert at approximately 75%)
Units with serial numbers of H20000 and greater: 83.5%±1.5%
(The audible alarm will alert at approximately 75%. At less than 60%, the red "service required" light will activate.)
-40°C to 70°C, humidity range of 10% to 100%, including condensation
Class II Equipment Double Insulated
TUV
IEC 601-1+A1+A2
EN 60601+A1+A2
ISO 8359:1996
DEVILBISS 5-LITER SERIES
515DS, 515DZ
1 to 5 LPM
5 LPM
8.5 psig (58.6 kPa)
115 V, 60 Hz, 4.2 Amp
97-127 V~, 60 Hz
1-5 LPM=93%±3%
Across the voltage range:
No degradation of performance
Tested at nominal voltage only:
No degradation in performance
Not recommended/Not tested
No degradation in performance across
the operating voltage range
(tested at sea level).
400 Watts Average
52 lbs. (23.5 Kilograms)
52.5 dbA overall average**
27.75"H x 16"W x 14"D (70.5 x 40.6 x 35.6 cm)
44 psig±3psig (303 kPa±21 kPa)
Time Cycle / Pressure Swing
Type B Applied Part
CSA
CAN/CSA-C22.2 No. 601.1-M90
IEC 601-1-2
515KS, 515KZ, 515NS
220-230 V~, 50 Hz, 1.7 Amp
230 V~, 60 Hz, 1.9 Amp
187-255 V~, 50 Hz
218-253 V~, 60 Hz
Across the voltage range:
No degradation in performance
Tested at 230V/50Hz only:
~4% oxygen perfomance drop at 5 LPM
Not recommended/Not tested
~4% oxygen performance drop at 5 LPM
and 35°C condition. No degradation in
performance across the operating voltage
range at other liter flows (tested at 670 M).
230v / 50 Hz - 365 Watts Average
230v / 60 Hz - 435 Watts Average
53 lbs. (24.5 Kilograms)
50.5 dbA overall average (50Hz)
TUV
IEC 601-1+A1+A2
EN 60601+A1+A2
ISO 8359:1996
(515KZ excluded)