8. DO ( Dissolved Oxygen )
MEASURING and CALIBRATION
PROCEDURE
8-1 Dissolved Oxygen measurement
1) Prepare the Oxygen Probe ( optional, DOPB-11 ),
install the " Probe Plug " ( 8-1, Fig. 5 ) into the
" DO Socket " ( 3-14, Fig. 1 ).
2) Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
Select the Meter's mode to " do " ( Dissolved
Oxygen " measurement, refer to chapter 4, page 12.
25
Fig. 5