Lutron Electronics WA-2017SD Operation Manual page 24

Sd card real time datalogger ph, orp, cd, tds, do, salt meter
Hide thumbs Also See for WA-2017SD:
Table of Contents

Advertisement

2) Install the " Probe Plug " ( 6-1, Fig. 3 ) into the
" CD Socket " ( 3-15, Fig. 1 ).
3) Power on the meter by pressing " Power Button "
( 3-2, Fig. 1 ) once.
Select the Meter's measurement Mode to " Cd "
( Conductivity measurement )
4) Hold the " Probe Handle " ( 6-2, Fig. 3 ) by hand
and let the " Sensing head " ( 6-3, Fig. 3 )
immersed wholly into the measured solution. Shake
the probe to let the probe's internal air bubble drift
out from the sensing head.
Display will show the conductivity mS ( uS ) values.
5) Use the two fingers to press the " REC Button " ( 3-4,
Fig 1 ) and " HOLD Button " ( 3-3, Fig. 1 ) at the same
time. the display will show the following screen as
example, release the both fingers.
Cd
CAL
6) Press the " Enter Button " ( 3-4, Fig. 1 ), the measuring
value will present on both upper and lower Display.
Use "
Button " ( 3-5, Fig. 1 ), "
to adjust the up display value exact same as the standard
conductivity value. Press the " Enter Button " ( 3-4. Fig. 1 )
will save the calibration data and finish the calibration
procedures.
Button " ( 3-6, Fig. 1 )
22