To toggle between two possible values
If only two values are possible, you can toggle between these values. Toggling
works for the pairs "Off"/"On", "Auto"/"Manual", "Left"/"Right", etc.
► Click the value to change to the other value.
5.3
R&S NRX
In a measurement, the R&S NRX uses all sensor-dependent measurement func-
tions and displays the results. Thus, you can configure both the measurement
and the power sensor.
Setup
1. Connect the power sensor to the R&S NRX.
See
Chapter 3.6.3, "R&S NRX base
2. Connect the power sensor to the signal source.
See
Chapter 3.4, "Connecting to a
Starting a measurement
1. Preset the R&S NRX and the connected R&S power sensors.
a) Press the [Preset] key.
b) Tap "Preset".
All parameters are set to their defaults.
2. If measuring in zero-IF mode (RBW > 40 MHz), consider to zero the power
sensor:
Note: Turn off all measurement signals before zeroing. An active measure-
ment signal during zeroing causes an error.
a) Switch off the power of the signal source.
b) Press the [Zero] key of the R&S NRX.
c) Tap "Zero All Sensors".
3. Configure the measurement.
a) In the "Measurement Settings" dialog, select the "Measurement Type", for
example "Continuous Average".
b) Tap "Quick Setup" > "Auto Set".
4. Switch on the signal source.
Getting Started 1178.3705.02 ─ 08
unit",
on page 28.
DUT",
on page 17.
Operating concepts
R&S NRX
45