ABB HART FEP300 Commissioning Instruction page 35

Electromagnetic flowmeter
Table of Contents

Advertisement

Important
For additional information regarding operation of the LCD display, refer to chapter
"Operation".
For detailed descriptions of all menus and parameters, see chapter "Parameterization" of
the operating instruction.
CI/FEP300/FEH300-EN
ProcessMaster / HygienicMaster
In this menu, the status of the current output can
be set in case of alarm (low or high alarm).
Press the ► key and selected the corresponding
parameter.
Press the ► key and set the value for the low
Alarm.
Setting range:
Factory setting:
Press the ► key and set the value for the high
Alarm.
Setting range:
Factory setting:
Press the ► to set the system zero. Valve must
be closed. The sensor must be filled completely.
The fluid must be at standstill. Press "OK" to start
the automatic adjust.
Once you have set all the parameters, you return
to the "Easy Set-up" menu. The most important
parameters are now set.
Press the ◄ key to "Exit" and return to the
process display.
Startup Operation
3.5 ... 3.6 mA
3.5 mA
21 ... 23 mA
21.8 mA
EN - 33