I N S T A L L A T I O N
5. INSTALL SONIC TRANSDUCER
WARNING: The Sonic Transducer is MR CONDITIONAL and
must be kept at least 3' away from the bore at all times .
Exercise extreme caution when handling it until it is
mounted .
5.1 Mounting Option 1 .
The most common area to mount the transducer is on
the floor near the front left or right side of the magnet.
This should be on the opposite side of the injector (Fig .
7). Wipe floor clean, remove sticker tabs from bottom of
transducer and place where desired .
5.1 Mounting Option 2 . Remove the lower side panel on
one side and mount the transducer under the magnet
(Fig. 8). Wipe floor clean, remove sticker tabs from bottom
of transducer and place where desired . Feed Pneumatic
Tubing through opening in machine base (Fig . 9) .
(Fig. 10)
(Fig. 12)
12
(Fig. 7)
5.2 Connect the 45' cable to the BNC port on the transducer (Fig .
10) . Pull any excess cable back toward the penetration panel leaving
enough slack if the transducer needs to be adjusted . Ensure that
there are no coils in cable . Coils are NOT MR SAFE . Collect the cable
slack at the penetration panel, coil it, flatten and zip-tie (Fig. 11). Stuff
as much excess cable as possible back into conduit .
5.3 Insert the male end of the 9' Pneumatic Tubing into the end port
of the transducer . (Fig . 12) . Make sure it is fully inserted . If mounted
inside MRI machine, carefully close panel, leaving enough space for
tubing (Fig . 13) .
(Fig. 13)
(Fig. 8)
(Fig. 9)
(Fig. 11)
Pneumatic Tubing
IM-1505 Rev. D