Leakage Testing; Pressure Transducer Verification - GE DINAMAP ProCare 100 Service Manual

Hide thumbs Also See for DINAMAP ProCare 100:
Table of Contents

Advertisement

4.6.2 Leakage Testing

4.6.3 Pressure Transducer Verification

4-12
Calibration & Maintenance: 4.6 ProCare Patient Monitor Parameter Tests
Note: To enter service mode press and hold the CYCLE button while
pressing the ON/OFF button
1. Turn the Monitor ON and enter Service Mode.
2. Press CYCLE button and the min display should change to a 1
3. Close the valve on the inflation bulb
4. Set the NIBP Analyzer to "Leak test" function
5. Press the "Pump" button on the NIBP Analyzer
6. After the system inflates to 200mmHg, wait for the "Wait" on the NIBP
analyzer message to be cleared
7. Press the "Start" button on the NIBP Analyzer to begin the leakage test
8. After 60 seconds the NIBP Analyzer will display the leakage rate
9. Record and verify the leakage rate.
10. Turn the ProCare Monitor off
Note: To enter service mode press and hold the CYCLE button while
pressing the ON/OFF button
1. Turn the Monitor ON and enter Service Mode.
2. The HISTORY LED should display 0.
3. Set NIBP Analyzer to "Manometer" function and press "Zero Pressure"
4. Press CYCLE button, HISTORY display should change to a 1.
5. Use the inflation bulb to inflate the cuff, hose and pressure indicator setup
to 200mmHg
6. Record and verify the pressure reading on the top LED display
(SYSTOLIC)
7. Record and verify the pressure reading on the bottom LED display
(DIASTOLIC)
8. Use the valve on the bulb to reduce pressure to 150mmHg
9. Record and verify the pressure reading on the top LED display
(SYSTOLIC)
10. Record and verify the pressure reading on the bottom LED display
(DIASTOLIC)
11. Use the valve on the bulb to reduce pressure to 100mmHg
ProCare Patient Monitor Service Manual
2009381-001