EMC INFORMATION

Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The MRIaudio Sound System is intended for use in the electromagnetic environment specified below. The customer or the
user of the MRIaudio Sound System should assure that it is used in such an environment .
EMISSIONS TEST & STANDARD
Radiated RF Emissions
CISPR 11
Conducted RF EMISSIONS
CISPR 11
Harmonic distortion
IEC 61000-3-2
Voltage fluctuations and flicker
IEC 61000-3-3
IMMUNITY TEST
AND STANDARD
Electrostatic Discharge
IEC 61000-4-2
0 % UT; 0,5 cycle, @
0°, 45°, 90°, 135°, 180°,
Voltage dips
225°, 270° and 315°
IEC 61000-4-11
70% UT; 25/30 cycles,
Voltage interruptions
0 % UT; 250/300 cycle
IEC 61000-4-11
Rated power freq .
magnetic fields
IEC 61000-4-8
Electrical fast
transients/burst
IEC 61000-4-4
Surges (Line-to-line)
IEC 61000-4-5
Surges (Line-to-
ground)
± 0 .5 kV, ±1 kV, ±2 kV
IEC 61000-4-5
20
COMPLIANCE
Group 1
Class A
Class A
Complies
IEC 60601 TEST
COMPLIANCE LEVEL
LEVEL
± 8 kV contact
± 8 kV contact
± 2 kV, ±4 kV, ±8 kV,
±15 kV air
0 % UT; 0,5 cycle, @
0°, 45°, 90°, 135°, 180°,
225°, 270° and 315°
0% UT; 1 cycle and
0% UT; 1 cycle and
70% UT; 25/30 cycles,
@ 0°
0 % UT; 250/300 cycle
30 A/m
50 Hz or 60 Hz
50 Hz or 60 Hz
± 2 kV for power
± 2 kV for power
supply lines
± 1 kV for input/
± 1 kV for input/
output lines
output lines
100 kHz repetition
100 kHz repetition
frequency
± 0 .5 kV, ±1 kV
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
The MRIaudio Sound System uses RF energy only for its internal
function . Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment .
The MRIaudio Sound System is suitable for use in all establishments
other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes .
ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Floors should be wood, concrete or ceramic tile .
If floors are covered with synthetic material, the
±15 kV air
relative humidity should be at least 30% .
Main power quality should be that of a typical
commercial or hospital environment . If the operator
requires continued operation during power
interruptions, it is recommended that the system be
powered from an uninterrupted power supply or a
@ 0°
battery .
Power frequency magnetic fields should be at levels
30 A/m
characteristic of a typical location in a typical
commercial or hospital environment .
supply lines
Main power quality should be that of a typical
commercial or hospital environment
frequency
± 1 kV
Main power quality should be that of a typical
commercial or hospital environment .
± 2 kV
IM-1505 Rev. D