Barocel 7025M Gauge
INSTRUCTION MANUAL
W601100880_A
Original Instructions

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Barocel 7025M and is the answer not in the manual?

Questions and answers

Summary of Contents for Edwards Barocel 7025M

  • Page 1 Barocel 7025M Gauge INSTRUCTION MANUAL W601100880_A Original Instructions...
  • Page 2 Product warranty and limit of liability are dealt with in Edwards standard terms and conditions of sale or negotiated contract under which this document is supplied. You must use the Barocel 7025M as described in this manual. Read this manual before you install, operate and maintain the Barocel 7025M.
  • Page 3 Burgess Hill West Sussex RH15 9TW The following product List products here including – • Barocel 7025m Series Capacitance Manometer W6011* Barocel 7025 Series Capacitance Manometer W6021* • • Barocel 7045 Series Capacitance Manometer W6032* • Barocel 7100 Series Capacitance Manometer W6033*...
  • Page 4 Additional Legislation and Compliance Information EU RoHS Directive: Material Exemption Information This product is compliant with no Annex III or IV Exemptions EU REACH Regulation Compliance This product is a complex article which is not designed for intentional substance release. To the best of our knowledge the materials used comply with the requirements of REACH.
  • Page 5: Table Of Contents

    W601100880_A - Contents Contents Introduction Product identification ....................7 Intended use ........................7 Function ......................... 7 Before using this gauge ....................7 Safety symbols ....................... 8 Safety precautions ......................8 Liability and warranty ....................8 Technical data Dimensions ........................11 Analogue measurement signal versus pressure ............
  • Page 6 W601100880_A - Contents This page has been intentionally left blank. Page 4...
  • Page 7 W601100880_A - List of Figures List of Figures Figure 1: Dimensions (mm) ....................11 Figure 2: Analogue measurement signal versus pressure ............12 Figure 3: Connect the gauge to the system ................15 Figure 4: FCC69 8-pin sensor cable ..................16 Figure 5: D-Sub 9-pin sensor cable ..................
  • Page 8 W601100880_A - List of Tables List of Tables Table 1: General ........................9 Table 2: Output signal ......................9 Table 3: Gauge identification ....................9 Table 4: Electrical ........................9 Table 5: Materials exposed to vacuum .................. 10 Table 6: Environmental ......................10 Page 6...
  • Page 9: Introduction

    Product identification Specify the information given on the gauge nameplate in all communications with Edwards. Intended use The Barocel 7025M gauge is intended for absolute pressure measurement of gases in their respective pressure ranges, see Installation on page 13. The unit is clean room compliant and double protected against contamination.
  • Page 10: Safety Symbols

    Ensure all other users are aware of the safety instructions. Liability and warranty Edwards accepts no liability and the warranty becomes non applicable if the end user or third parties: Ignore the information in this document ...
  • Page 11: Technical Data

    Protective Extra-Low Voltage (PELV) and Limited Power Source (LPS), Class 2. The connection to the gauge has to be fused. Edwards controllers meet this requirement. Table 4 Electrical Supply voltage at the gauge +13 to +32 V d.c.
  • Page 12: Table 5 Materials Exposed To Vacuum

    W601100880_A - Technical data Table 4 Electrical (continued) Electrical connection FCC68/RJ45, 8 pin socket or D-Sub, 9-pin, male Sensor cable 8-pin or 9-pin plus shielding § Cable length  100 m (0.14 mm² conductor) For longer cables, larger conductor cross-sections are required. Grounding concept Vacuum flange - signal common Electrical connection...
  • Page 13: Dimensions

    W601100880_A - Technical data Dimensions Figure 1 Dimensions (mm) Page 11...
  • Page 14: Analogue Measurement Signal Versus Pressure

    W601100880_A - Technical data Analogue measurement signal versus pressure Figure 2 Analogue measurement signal versus pressure p = (U / 10 V) × p (F.S.) Conversion Torr to Pascal Torr mbar 1.00 1013.25/760 = 1.3332… 101325 / 760 = 133.3224… Source: NPL (National Physical Laboratory) Guide to the Measurement of Pressure and Vacuum, ISBN 0904457x /1998 Example:...
  • Page 15: Installation

    W601100880_A - Installation Installation WARNING: NOT FAIL-SAFE DESIGNED Risk of death. Do not use the Barocel 7025M for safety critical applications. The gauge is not intended to be fail-safe. CAUTION: IMPACT DAMAGE Risk of damage to equipment. The ceramic sensor can be damaged by impacts. Do not drop the gauge.
  • Page 16 W601100880_A - Installation 1. Position the gauge to prevent vibrations. Note: The gauge can be installed in any orientation. We recommend you install the gauge in a horizontal to upright position and use a seal with a centring ring and filter to prevent the influx of condensates and particles in the measurement chamber.
  • Page 17: Electrical Connection

    Voltage (PELV) and Limited Power Source (LPS), Class 2. The connection to the gauge has to be fused. Edwards controllers meet this requirement. Make sure the vacuum connection is correct. Ground loops, differences of potential, or EMC problems can affect the measurement signal.
  • Page 18: Fcc68, 8-Pin Connector

    W601100880_A - Installation Use differential measurement input (signal common and supply common wired  separately). Potential difference between the supply common and the housing 16 V (over-  voltage protection). FCC68, 8-pin connector 1. If a sensor cable is not available, one can be made. Refer to Figure 4 on page 16.
  • Page 19: D-Sub, 9-Pin Connector

    W601100880_A - Installation D-Sub, 9-pin connector 1. If a sensor cable is not available, one can be made. Refer to Figure 5 on page 17. Connect the sensor cable. Figure 5 D-Sub 9-pin sensor cable Signal output (measurement signal) N.C. N.C.
  • Page 20: Operation

    W601100880_A - Operation Operation Put the gauge into operation. If you use an Edwards controller, specify the measurement range. Warm-up time: approximately 1 minute To zero the gauge The gauge is factory calibrated in an upright orientation. It requires no maintenance.
  • Page 21 W601100880_A - Operation 2. Briefly press the <ZERO> button with a pin (maximum ø1.1 mm). The zero adjustment runs automatically (duration 8 seconds). Press the pin briefly Note: After zero adjustment, the gauge automatically returns to measurement mode. Page 19...
  • Page 22: Uninstallation

    W601100880_A - Uninstallation Uninstallation WARNING: CONTAMINATED PARTS Risk to health and the environment from contaminated parts. Before you start work, find out if any parts are contaminated. Obey the relevant regulations and follow the necessary precautions when handling contaminated parts. CAUTION: IMPACT DAMAGE Risk of damage to equipment.
  • Page 23: Maintenance And Repair

    The warranty does not include gauge failures that result from contamination. We recommend that you check the zero at regular intervals. Edwards accepts no liability and the warranty becomes non applicable if the end-user or third parties do repair work.
  • Page 24: Disposal

    W601100880_A - Disposal Disposal WARNING: CONTAMINATED PARTS Risk to health and the environment from contaminated parts. Before you dispose of the gauge, find out if any parts are contaminated. Obey the relevant regulations and observe the necessary precautions when handling contaminated parts. WARNING: ENVIRONMENTAL POLLUTANT Risk to environment from substances or parts.
  • Page 25 This page has been intentionally left blank.
  • Page 26 This page has been intentionally left blank.