Barocel 7025 Gauge
INSTRUCTION MANUAL
W602100880_A
Original Instructions

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Barocel 7025 and is the answer not in the manual?

Questions and answers

Summary of Contents for Edwards Barocel 7025

  • Page 1 Barocel 7025 Gauge INSTRUCTION MANUAL W602100880_A Original Instructions...
  • Page 2 Product warranty and limit of liability are dealt with in Edwards standard terms and conditions of sale or negotiated contract under which this document is supplied. You must use the Barocel 7025 as described in this manual. Read this manual before you install, operate and maintain the Barocel 7025.
  • Page 3 West Sussex RH15 9TW The following product List products here including – • Barocel 7025m Series Capacitance Manometer W6011* Barocel 7025 Series Capacitance Manometer W6021* • • Barocel 7045 Series Capacitance Manometer W6032* • Barocel 7100 Series Capacitance Manometer W6033*...
  • Page 4 Additional Legislation and Compliance Information EU RoHS Directive: Material Exemption Information This product is compliant with no Annex III or IV Exemptions EU REACH Regulation Compliance This product is a complex article which is not designed for intentional substance release. To the best of our knowledge the materials used comply with the requirements of REACH.
  • Page 5: Table Of Contents

    W602100880_A - Contents Contents Introduction Product identification ....................7 Intended use ........................7 Function ......................... 7 Before using this gauge ....................7 Safety symbols ....................... 7 Safety precautions ......................7 Liability and warranty ....................8 Technical data Dimensions ........................11 Analogue measurement signal versus pressure ............
  • Page 6 W602100880_A - Contents This page has been intentionally left blank. Page 4...
  • Page 7 W602100880_A - List of Figures List of Figures Figure 1: Dimensions ......................11 Figure 2: Analogue measurement signal versus pressure ............12 Figure 3: Connect the gauge to the system ................14 Figure 4: Sensor cable ......................15 Figure 5: Gauge display ......................16 Figure 6: Measurement signal (Pressure p) ................
  • Page 8 W602100880_A - List of Tables List of Tables Table 1: General ........................9 Table 2: Output signal ......................9 Table 3: Gauge identification ....................9 Table 4: Switching functions ....................9 Table 5: Electrical ........................10 Table 6: Materials exposed to vacuum .................. 10 Table 7: Environmental ......................
  • Page 9: Introduction

    Product identification Specify the information given on the gauge nameplate in all communications with Edwards. Intended use The Barocel 7025 gauge is intended for absolute pressure measurement of gases in their respective pressure ranges, see Installation on page 13. The unit is clean room compliant and double protected against contamination.
  • Page 10: Liability And Warranty

    Ensure all other users are aware of the safety instructions. Liability and warranty Edwards accepts no liability and the warranty becomes non applicable if the end user or third parties: Ignore the information in this document ...
  • Page 11: Technical Data

    W602100880_A - Technical data Technical data Table 1 General Measurement range Accuracy 1.0 Torr/mbar (F.S.) 0.20% of reading 0.1 Torr/mbar (F.S.) 0.50% of reading Temperature effect on zero 10.0 Torr/mbar (F.S.) 0.0050% F.S./°C 0.1 Torr/mbar (F.S.) 0.020% F.S./°C Temperature effect on span 1.0 Torr/mbar (F.S.) 0.01% of reading/°C 0.1 Torr/mbar (F.S.)
  • Page 12: Table 5 Electrical

    W602100880_A - Technical data WARNING: HIGH VOLTAGE Risk of electrical shock. The gauge can only be connected to power supplies, instruments or control devices that comply with the requirements of a grounded Protective Extra-Low Voltage (PELV) and Limited Power Source (LPS), Class 2. The connection to the gauge has to be fused.
  • Page 13: Dimensions

    W602100880_A - Technical data Table 7 Environmental Admissible temperatures Storage -40 °C to +65 °C Operation +5 °C to +50 °C Bakeout (not in operation) 110 °C at the flange 80% at temperatures Relative humidity +31 °C decreasing to 50% at +40 °C Indoors only, altitude up to 2000 metres NN Degree of protection IP 30...
  • Page 14: Analogue Measurement Signal Versus Pressure

    W602100880_A - Technical data Analogue measurement signal versus pressure Figure 2 Analogue measurement signal versus pressure p = (U / 10 V) × p (F.S.) Conversion Torr to Pascal Torr mbar 1.00 1013.25 / 760 = 1.3332… 101325 / 760 = 133.3224… Example: Gauge with 10 Torr F.S.
  • Page 15: Installation

    W602100880_A - Installation Installation WARNING: NOT FAIL-SAFE DESIGNED Risk of death. Do not use the Barocel 7025 for safety critical applications. The gauge is not intended to be fail-safe. CAUTION: IMPACT DAMAGE Risk of damage to equipment. The ceramic sensor can be damaged by impacts. Do not drop the gauge.
  • Page 16 W602100880_A - Installation 1. Position the gauge to prevent vibrations. Note: The gauge can be installed in any orientation. We recommend you install the gauge in a horizontal to upright position and use a seal with a centring ring and filter to prevent the influx of condensates and particles in the measurement chamber.
  • Page 17: Electrical Connection

    Voltage (PELV) and Limited Power Source (LPS), Class 2. The connection to the gauge has to be fused. Edwards controllers meet this requirement. Make sure the vacuum connection is correct. Ground loops, differences of potential, or EMC problems can affect the measurement signal.
  • Page 18: Figure 4 Sensor Cable

    W602100880_A - Installation 1. If a sensor cable is not available, one can be made. Refer to Figure 4 on page 16. Figure 4 Sensor cable 1, 4 Relay SP1, closing contact Signal output (measurement signal) or thresholds SP1/2 Supply common, ground 7, 11 Supply 8, 9...
  • Page 19: Operation

    W602100880_A - Operation Operation Put the gauge into operation. If you use an Edwards controller, specify the measurement range and the unit of measurement. Warm-up time: For general purpose reading (within specifications) > 0.25 hours  For zero adjustment and precision measurement >...
  • Page 20: Zero> Adjustment

    The <ZERO> button on the gauge  The RS232C interface, see Operation on page 17  An Edwards vacuum gauge controller.  To adjust the zero: 1. Evacuate the gauge to a pressure according to the table below: F.S. Recommended final pressure for zero adjustment 1100 mbar <...
  • Page 21: Zero> Adjustment With Ramp Function

    W602100880_A - Operation Press the pin briefly Note: After zero adjustment, the gauge automatically returns to measurement mode. The <RUN> LED flashes if: At final pressure, the signal output is negative (< -20 mV)  The zero adjustment has failed. ...
  • Page 22: Switching Functions

    W602100880_A - Operation Keep the pin depressed 3. Push the <ZERO> button again: Fine adjustment within 0 to 3 seconds: The zero adjustment value changes by one unit (push <ZERO> button in 1 second intervals) Change of direction within 3 to 5 The zero adjustment changes its direction (the seconds: flashing frequency of the <RUN>...
  • Page 23: Adjusting The Setpoints

    W602100880_A - Operation Figure 6 Measurement signal (Pressure p) Setpoint Hysteresis Threshold value Adjusting the setpoints CAUTION: DAMAGE TO THE EQUIPMENT Risk of malfunction. If processes are controlled by the signal output, by pushing the <SP> button, the measurement signal is suppressed and the corresponding threshold value is output instead.
  • Page 24 W602100880_A - Operation Press the button 2. To change the threshold value, push the <ZERO> button and keep it depressed. The threshold keeps changing from the current value (ramp) until the button is released or until you get to the limit of the setting range. Keep the pin depressed 3.
  • Page 25: Activating The Factory Setting (Factory Reset)

    W602100880_A - Operation Fine adjustment within The zero adjustment value changes by one unit. 0 to 3 seconds: Change of direction The zero adjustment changes its direction (the flashing within 3 to 5 seconds: frequency of the <RUN> LED changes briefly). Note: If the <ZERO>...
  • Page 26: Uninstallation

    W602100880_A - Uninstallation Uninstallation WARNING: CONTAMINATED PARTS Risk to health and environment from contaminated parts. Before you start work, find out if any parts are contaminated. Obey the relevant regulations and follow the necessary precautions when handling contaminated parts. CAUTION: IMPACT DAMAGE Risk of damage to equipment.
  • Page 27: Maintenance And Repair

    The warranty does not include gauge failures that result from contamination. We recommend that you check the zero at regular intervals. Edwards accepts no liability and the warranty becomes non applicable if the end user or third parties do repair work.
  • Page 28: Disposal

    W602100880_A - Disposal Disposal WARNING: CONTAMINATED PARTS Risk to health and the environment from contaminated parts. Before you dispose of the gauge, find out if any parts are contaminated. Obey the relevant regulations and observe the necessary precautions when handling contaminated parts. WARNING: ENVIRONMENTAL POLLUTANT Risk to environment from substances or parts.