battery displayed). Although the device will continue to measure
reliably, you should obtain replacement batteries.

Flat battery - replacement

When the batteries are flat, the battery symbol AM will flash as soon
as the device is switched on (flat battery displayed). You cannot
take any further measurements and must replace the batteries.
1. Open the battery compartment 6 at the back of the device.
2. Replace the batteries – ensure correct polarity as shown by the
symbols in the compartment.
3. To set date and time, follow the procedure described in Section
«2. Using the device for the first time».
The measurements stored in the memory are deleted when
the batteries are removed from the battery compartment
(e.g. when replacing batteries).

Which batteries and which procedure?

Use 4 new, long-life 1.5 V, size AAA alkaline batteries.
Do not use batteries beyond their date of expiry.
Remove batteries if the device is not going to be used for a
prolonged period.

8. Error messages

If an error occurs during the measurement, the measurement is
interrupted and an error message, e.g. «ERR 3», is displayed.
Error
Description
«ERR 1»
Signal too
weak
AR
«ERR 2»
Error signal
AQ-B
«ERR 3»
Abnormal cuff
pressure
AQ-C
Microlife BP B1 Classic
Potential cause and remedy
The pulse signals on the cuff are too
weak. Re-position the cuff and repeat
the measurement.*
During the measurement, error
signals were detected by the cuff,
caused for instance by movement or
muscle tension. Repeat the
measurement, keeping your arm still.
An adequate pressure cannot be
generated in the cuff. A leak may
have occurred. Check that the cuff is
correctly connected and is not too
loose. Replace the batteries if neces-
sary. Repeat the measurement.
Error
Description
«ERR 5» Abnormal
result
«HI»
Pulse or cuff
pressure too
high
«LO»
Pulse too low
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.

9. Safety, care, accuracy test and disposal

m
Safety and protection
 Follow instructions for use. This document provides important
product operation and safety information regarding this device.
Please read this document thoroughly before using the device
and keep for future reference.
 This device may only be used for the purposes described in
these instructions. The manufacturer cannot be held liable for
damage caused by incorrect application.
 This device comprises sensitive components and must be
treated with caution. Observe the storage and operating condi-
tions described in the «Technical specifications» section.
 The cuffs are sensitive and must be handled with care.
 Only pump up the cuff once fitted.
 Do not use this device if you think it is damaged or notice
anything unusual.
 Never open this device.
 Read the additional safety information provided within the indi-
vidual sections of this instruction manual.
 The measurement results given by this device is not a diag-
nosis. It is not replacing the need for the consultation of a physi-
cian, especially if not matching the patient's symptoms. Do not
rely on the measurement result only, always consider other
Potential cause and remedy
The measuring signals are inaccu-
rate and no result can therefore be
displayed. Read through the check-
list for performing reliable measure-
ments and then repeat the measure-
ment.*
The pressure in the cuff is too high
(over 299mmHg) OR the pulse is too
high (over 200 beats per minute).
Relax for 5 minutes and repeat the
measurement.*
The pulse is too low (less than 40
beats per minute). Repeat the
measurement.*