Defaults - Emerson Rosemount 1054B Instruction Manual

Microprocessor ph/orp analyzer
Table of Contents

Advertisement

MODEL 1054B pH/ORP
4.8 DEFAULTS. Display Mnemonic dFt. This item is
used to set the configuration of relays and output
default conditions during fault or hold status. See
Table 8-1 for a listing of the possible fault conditions
which can be diagnosed by the analyzer.
A. Relay 1 and 2. Display Mnemonic rL1 and rL2.
During a fault or hold condition the relays can be set
to activate on, deactivate oFF, or remain in the state
determined by the last process value non. See Table
4-3.
B. Current Output. Display Mnemonic Cur. The cur-
rent output is held at the last process value non or
goes to a specified value in mA cur during a fault or
hold condition.
C. Fault History. Display Mnemonic SHO. SELECT this
item will sequence the display through all faults
detected in most recent fault mode. Press the
SCROLL ( é ) key once for previous fault mode list.
Pressing ACCESS will clear SHO history.
Set menu
default
Set men AL1/AL2 setting
(dFt)
On
setting
rL1/rL2
on
Proc. det.
off
Proc. det.
non
Proc. det.
Proc. det.: Alarm state is determined by the process value
+
: Relay will activate
: Relay will not activate
Example: If you want the analyzer to activate relay 1 in hold mode during buffer calibration, set
AL1 to On in Section 4.2, and set rL1 to on.
28
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
TABLE 4-3. Relay States for Various Analyzer Conditions
and Alarm/Default Configurations
NORMAL
OFF
FLt
(Alarm 2
only)
4.8.1 Default Setup (dFt).
1.
Enter Set Mode by pressing the ACCESS key
twice.
SCROLL ( é ) until dFt appears on the display.
2.
3.
SELECT to move to the next menu level. rL1 will
display.
4.
SCROLL ( é ) then SELECT desired item.
5.
Display will show next item selection. SCROLL
( é ) and ENTER desired item.
6.
Repeat Steps 4 and 5 as required for other
default settings rL2 and Cur. If cur is selected for
Cur, press ENTER, then use the SCROLL ( é )
and SHIFT (ç) keys to enter the desired current
value for a fault or hold condition.
7.
Press the ENTER key to return to Set Menu.
ANALYZER CONDITION
HOLD
Set menu AL1/AL2 setting
On
OFF
FLt
(Alarm 2
only)
+
Proc. det.
SECTION 4.0
CONFIGURATION
FAULT
Set menu AL1/AL2 setting
On
OFF
FLt
(Alarm 2
only)
+
+
+
Proc. det.
+