Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The AT235 is intended f or use in the electromagnetic environment specif ied below. The customer or the user
of the AT235 should assure that it is used in such an environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical f ast
transient/burst
IEC61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
Power f requency
(50/60 Hz)
IEC 61000-4-8
Note: UT is the A.C. mains voltage prior to application of the test level.
D-0128073-C – 2022/05
AT23 5 - Instructions for Use - US
IEC 60601 Test
Level
+6 kV contact
+8 kV air
+2 kV f or power supply
lines
+1 kV f or input/output
lines
+1 kV dif f erential mode
+2 kV common mode
< 5% UT
(>95% dip in UT) f or 0.5
cycle
40% UT
(60% dip in UT) f or 5
cycles
70% UT
(30% dip in UT) f or 25
cycles
<5% UT
(>95% dip in UT) f or 5
sec
3 A/m
Compliance
+6 kV contact
+8 kV air
+2 kV f or power supply
lines
+1 kV f or input/output
lines
+1 kV dif f erential mode
+2 kV common mode
< 5% UT (>95% dip in
UT)
f or 0.5 cycle
40% UT (60% dip in
UT) f or
5 cycles
70% UT (30% dip in
UT) f or
25 cycles
<5% UT
3 A/m
Electromagnetic
Environment-Guidance
Floors should be wood, concrete
or ceramic tile. If f loors are
covered with synthetic material,
the relative humidity should be
greater than 30%.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of the AT235 requires
continued operation during
power mains interruptions, it is
recommended that the AT235 be
powered f rom an uninterruptable
power supply or its battery.
Power f requency magnetic f ields
should be at levels characteristic
of a typical location in a typical
commercial or residential
environment.