Probe Status - Interacoustics AT235 Instructions For Use Manual

Hide thumbs Also See for AT235:
Table of Contents

Advertisement

2.14 Probe Status

The probe status is indicated by the color of the light on the control panel, standard probe system and the
clinical probe system. Below, the colors and their meanings are explained:
Color
Control panel
Red
Blue
Green
Yellow
White
Blinking
No light
D-0128073-C – 2022/05
AT23 5 - Instructions for Use - US
Standard probe
Clinical probe
Status
Right ear is selected. Probe is out of
the ear.
Lef t ear is selected. Probe is out of
the ear.
Probe is in the ear and a seal is
maintained.
Probe is in the ear and is blocked,
leaking or too noisy.
The probe has just been attached.
Probe status is unknown. If the probe
light stays white in any other situation
the AT235 may need to be switched
of f and back on again to regain
proper probe status.
AT235 is pausing and/or an
interaction is expected. The AT235
will f or example remain blinking green
if the protocol has f inished testing and
the probe is still in the ear. The user
may pause the AT235 bef ore
inserting the probe resulting in blue or
red blinking.
AT235 is not monitoring the probe
status.
Page 17