Instructions for use PROPHYflex 4 Perio Kit – 1.011.9403 Refill PROPHYflex perio tip – 1.010.0287
6 Reprocessing steps in accordance with ISO 17664 | 6.4 Automated reprocessing
Consumables required:
▪ Cloths for wiping the medical device.
▶ Spray the disinfectant on a cloth, then wipe down the medical device and
allow the disinfectant to act according to the instructions of the
disinfectant manufacturer.
▶ Follow the instructions for use of the disinfectant.
6.3.4 Manual disinfection - internal
Manual internal disinfection of the PROPHYflex 4 perio tip is not applicable. For
effective reprocessing, the inside of the device must be subjected to automated
cleaning in a cleaning and disinfection unit in accordance with ISO 15883-1.
6.3.5 Manual drying
Manual drying of the PROPHYflex 4 perio tip is not applicable. For effective re-
processing, the inside of the device must be subjected to automated cleaning in
a cleaning and disinfection unit in accordance with ISO 15883-1.
6.4 Automated reprocessing
Incomplete disinfection.
Infection hazard.
▶ Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
▶ If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶ Remove the tool.
Never disinfect the medical device with chloride-containing products.
Malfunction and material damage.
▶ Only disinfect in the washer disinfector.
Never reprocess this medical device in an ultrasonic device.
Malfunction and material damage.
▶ Clean it in a washer disinfector only.
WARNING
WARNING
NOTICE
NOTICE
18 / 26