Table of Contents
loading
Need help?

Need help?

Do you have a question about the MULTIflex Coupling 465 RN and is the answer not in the manual?

Questions and answers

Summary of Contents for KaVo MULTIflex Coupling 465 RN

  • Page 1 Instructions for use MULTIflex Coupling 465 RN – 0.553.1600 MULTIflex LUX Coupling 465 LRN – 0.553.1550...
  • Page 2 Distribution: Manufacturer: KaVo Dental GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 88400 Biberach 88400 Biberach Germany Germany Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 Table of contents Table of contents 1 User instructions ....................... 2 Safety..........................2.1 Infection hazard ......................7 2.2 Electrical shock......................7 2.3 Use ..........................7 2.4 Technical condition ......................
  • Page 4: User Instructions

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
  • Page 5 Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 1 User instructions Information on the packaging Material number Serial number Manufacturer Manufacturing date Note: Please note accompanying documents Follow the electronic instructions for use...
  • Page 6 Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 1 User instructions CAUTION In cases which – if not prevented – can lead to minor or moderate injury. CAUTION In cases which – if not prevented – can lead to material damage.
  • Page 7: Safety

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
  • Page 8: Accessories And Combination With Other Equipment

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 2 Safety | 2.5 Accessories and combination with other equipment To ensure optimum function and to prevent property damage, please comply with the following instructions: 4 Service the medical device regularly with care products and systems as de- scribed in the instructions for use.
  • Page 9: Disposal

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 2 Safety | 2.8 Disposal 2.8 Disposal This product is subject to the EC directive governing waste electrical and elec- tronic devices and must be submitted to special disposal in Europe.
  • Page 10: Product Description

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 3 Product description | 3.1 Intended use 3 Product description MULTIflex Coupling 465 RN (Mat. no. 0.553.1600) MULTIflex LUX Coupling 465 LRN (Mat. no. 0.553.1550)
  • Page 11 Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 3 Product description | 3.1 Intended use ▪ to protect him or herself, the patient and third parties from hazards ▪ to prevent contamination from the product...
  • Page 12: Technical Specifications

    4 1065 / 1060: If the KaVo MULTIflex LUX coupling is used on these units, have a technician adjust the voltage. 4 E80 / E70 / E50 / 1058 / 1080 / 1066 / 1062: The KaVo MULTIflex LUX coupling can be used on these units with any further measures being re- quired.
  • Page 13: Transportation And Storage Conditions

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 3 Product description | 3.3 Transportation and storage conditions 3.3 Transportation and storage conditions CAUTION Startup after refrigerated storage. Malfunction. 4 Prior to startup, strongly refrigerated products must be allowed to warm up to a temperature of 20 °C to 25 °C (68 °F to 77 °F).
  • Page 14: Startup And Shut Down

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 4 Startup and shut down 4 Startup and shut down WARNING Hazard from contaminated products. Infection hazard to the dentist and patient. 4 Prior to initial startup and after each use, reprocess the product and ac- cessories.
  • Page 15: Operation

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 5 Operation | 5.1 Connecting the coupling to the hose 5 Operation NOTE At the beginning of each workday, the water-conducting systems should be...
  • Page 16: Pulling The Handpiece Off

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 5 Operation | 5.4 Pulling the handpiece off 4 Accurately attach the MULTIflex (LUX) instrument to the MULTIflex coupling and push it to the rear until the coupling audibly locks into the medical device.
  • Page 17: Checking For Malfunctions And Troubleshooting

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions before initial start-up 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions before initial start-up CAUTION Danger from hot lamp.
  • Page 18: Replacing The O-Rings

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Take lamp from the mount. 4 Insert the new lamp very carefully. 6.2.2 Replacing the O-rings CAUTION Missing or damaged O-rings.
  • Page 19: Reprocessing Steps In Accordance With Iso 17664

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparations at the site of use...
  • Page 20: Manual Internal And External Disinfection

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing For effective reprocessing, automated internal and external cleaning as well as automated internal and external disinfection with a cleaning and disinfection unit in accordance with EN ISO 15883-1 is required.
  • Page 21: Preparation For Automated Internal And External Cleaning As Well As Internal And Ex

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing CAUTION Never reprocess this medical device in an ultrasonic cleaner. Malfunction and material damage.
  • Page 22: Care Products And Systems - Servicing

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Care products and systems - Servicing NOTE Please comply with the instructions for use of the washer disinfector.
  • Page 23 Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.8 Storage NOTE Comply with the expiry date of the sterilized items. 23 / 28...
  • Page 24: Optional Aids And Consumables

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. Wrench 0.411.1563 Spare washer 465 RN 0.553.1872...
  • Page 25: Terms And Conditions Of Warranty

    Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...