Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 1 User instructions 1 User instructions Dear user, Congratulations on purchasing this KaVo quality product. By following the notes below you will be able to work smoothly, economically and safely.
Page 5
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 1 User instructions Information on the packaging Material number Serial number Manufacturer Manufacturing date Note: Please note accompanying documents Follow the electronic instructions for use...
Page 6
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 1 User instructions CAUTION In cases which – if not prevented – can lead to minor or moderate injury. CAUTION In cases which – if not prevented – can lead to material damage.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 2 Safety | 2.1 Infection hazard 2 Safety NOTE All serious events occurring in relation to the product must be reported to the manufacturer and the competent authority of the member state, in which the user and/or patient resides.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 2 Safety | 2.5 Accessories and combination with other equipment To ensure optimum function and to prevent property damage, please comply with the following instructions: 4 Service the medical device regularly with care products and systems as de- scribed in the instructions for use.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 2 Safety | 2.8 Disposal 2.8 Disposal This product is subject to the EC directive governing waste electrical and elec- tronic devices and must be submitted to special disposal in Europe.
Page 11
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 3 Product description | 3.1 Intended use ▪ to protect him or herself, the patient and third parties from hazards ▪ to prevent contamination from the product...
4 1065 / 1060: If the KaVo MULTIflex LUX coupling is used on these units, have a technician adjust the voltage. 4 E80 / E70 / E50 / 1058 / 1080 / 1066 / 1062: The KaVo MULTIflex LUX coupling can be used on these units with any further measures being re- quired.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 3 Product description | 3.3 Transportation and storage conditions 3.3 Transportation and storage conditions CAUTION Startup after refrigerated storage. Malfunction. 4 Prior to startup, strongly refrigerated products must be allowed to warm up to a temperature of 20 °C to 25 °C (68 °F to 77 °F).
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 4 Startup and shut down 4 Startup and shut down WARNING Hazard from contaminated products. Infection hazard to the dentist and patient. 4 Prior to initial startup and after each use, reprocess the product and ac- cessories.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 5 Operation | 5.1 Connecting the coupling to the hose 5 Operation NOTE At the beginning of each workday, the water-conducting systems should be...
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 5 Operation | 5.4 Pulling the handpiece off 4 Accurately attach the MULTIflex (LUX) instrument to the MULTIflex coupling and push it to the rear until the coupling audibly locks into the medical device.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 6 Checking for malfunctions and troubleshooting | 6.1 Checking for malfunctions before initial start-up 6 Checking for malfunctions and troubleshooting 6.1 Checking for malfunctions before initial start-up CAUTION Danger from hot lamp.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 6 Checking for malfunctions and troubleshooting | 6.2 Troubleshooting 4 Take lamp from the mount. 4 Insert the new lamp very carefully. 6.2.2 Replacing the O-rings CAUTION Missing or damaged O-rings.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.1 Preparations at the site of use 7 Reprocessing steps in accordance with ISO 17664 7.1 Preparations at the site of use...
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing For effective reprocessing, automated internal and external cleaning as well as automated internal and external disinfection with a cleaning and disinfection unit in accordance with EN ISO 15883-1 is required.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.4 Automated reprocessing CAUTION Never reprocess this medical device in an ultrasonic cleaner. Malfunction and material damage.
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.5 Care products and systems - Servicing NOTE Please comply with the instructions for use of the washer disinfector.
Page 23
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 7 Reprocessing steps in accordance with ISO 17664 | 7.8 Storage NOTE Comply with the expiry date of the sterilized items. 23 / 28...
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 8 Optional aids and consumables 8 Optional aids and consumables Available from specialised dental dealers. Material summary Mat. no. Wrench 0.411.1563 Spare washer 465 RN 0.553.1872...
Instructions for use MULTIflex Coupling 465 RN – 0.553.1600, MULTIflex LUX Coupling 465 LRN – 0.553.1550 9 Terms and conditions of warranty 9 Terms and conditions of warranty This KaVo medical device is subject to the following warranty conditions: KaVo grants the end customer a warranty of proper function and guarantees...
Need help?
Do you have a question about the MULTIflex Coupling 465 RN and is the answer not in the manual?
Questions and answers