4 Startup and shut down

Hazard from contaminated products.
Infection hazard for care provider and patient.
▶ Prior to initial commissioning and after each use, reprocess the product and
accessories.
Dispose of the product in appropriate manner.
Infection hazard.
▶ Before disposal, reprocess and sterilise the product and accessories appro-
priately.
▶ Dispose of the perio tip like injection needles in a closed container.
See also:
2 6 Reprocessing steps in accordance with ISO 17664, Page 16

4.1 Inserting the adaptor into the PROPHYflex

Adapter falling off during the treatment.
The adapter falling off is a hazard for patient and user.
Visual inspection each time the adapter is inserted.
Before commencing the treatment, check that there is no gap between the ad-
apter and the gripping sleeve.
▶ Insert the adapter in the gripping sleeve and use the wrench for the can-
nula (Mat. no. 3.004.6351) to tighten it in clockwise direction while hold-
ing on to the gripping sleeve.

4.2 Inserting the PROPHYflex perio tip into the wrench

▶ Slide the PROPHYflex perio tip into the wrench until it locks into place.
4 Startup and shut down | 4.1 Inserting the adaptor into the PROPHYflex
WARNING
WARNING
CAUTION
11 / 26