Greenlee CMPL-100 Instruction Manual page 19

Clamp-on meter
Table of Contents

Advertisement

Available languages

Available languages

Accuracy
See the Specifications section for operating conditions.
Accuracy is specified as follows:
decimal place) at 18 C to 28 C (64 F to 82 F), 0% to 80% relative humidity.
Accuracy Table
Value
True Power
600.0 kW
Apparent Power
600.0 kVA
0.3 to 1.0
(
= 72.5 to 0 )
Power Factor
0.0 to 0.3
(
= 90 to 72.5 )
DCA
ACA††
Peak Current*
DCV**
ACV** ††
Peak Voltage***
Continuity
Resistance
9.999 k
Frequency†
999.9 Hz
* Peak detect acquisition time
** Input impedance is 1 M .
*** Peak detect acquisition time
† Sensitivity is 10 V or 10 A.
†† AC-coupled, true RMS measurement,
additional error for crest factors > 1:
Crest Factor
Error (% of reading)
1 to 3
3 to 5
5 to 7
(a percentage of the reading + the value of the rightmost
Range
Accuracy
(2% + 0.5 kW)
(2% + 0.5 kVA)
3
Not
Specified
999.9 A
(2% + 0.5 A)
999.9 A
(2% + 1.0 A)
999.9 A
(2.5% + 1.0 A)
600.0 V
(0.75% + 0.2 V)
600.0 V
(1.2% + 1.0 V)
600.0 V
(2.5% + 1.0 V)
999.9
(1% + 0.5 )
(1% + 0.005 k )
(0.5% + 0.5 Hz)
100 ms
200 ms
0.5%
3%
6%
CMPL-100
Frequency Range
40 to 400 Hz
40 to 400 Hz
10 to 60 Hz
N/A
40 to 400 Hz
40 to 400 Hz
N/A
40 to 400 Hz
40 to 400 Hz
N/A
N/A
0 to 1 kHz
19