Advantech POC-S198-25F-ACE User Manual page 32

Table of Contents

Advertisement

Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The model POC-S198 is intended for use in the electromagnetic environment specified
below. The customer or the user of the model POC-S198 should assure that it is used in
such an environment.
Immunity
test
Conducted
RF
IEC 61000-4-
6
Radiated RF
IEC 61000-4-
3
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption
and reflection from structures, objects and people.
POC-S198 User Manual
IEC 60601 Test
Compliance
Level
Level
3 Vrms
150 kHz to 80
MHz
Vrms
3 V/m
V/m
80 MHz to 2,5
GHz
Electromagnetic Environmental
Guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the model
POC-S198, including cables, than
the recommended separation dis-
tance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Dis-
tance
d = 1,2
P
d = 1,2
P
80 MHz to 800 MHz
d = 2,3
P
800 MHz to 2,5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmit-
ter manufacturer and d is the rec-
ommended separation distance in
meters (m).
Field strengths from fixed RF trans-
mitters, as determined by an elec-
tromagnetic site survey,
less than the compliance level in
each frequency range.
Interference may occur in the vicin-
ity of equipment marked with the fol-
lowing symbol:
18