Advantech POC-S198-25F-ACE User Manual page 31

Table of Contents

Advertisement

Guidance and Manufacturer's Declaration – Electromagnetic Immunity
POC-S198 is intended for use in the electromagnetic environment specified below. The
customer or the user of the model POC-S198 should assure that it is used in such an envi-
ronment.
IEC 60601 Test
Immunity Test
Level
Electrostatic dis-
±6 kV contact
charge (ESD)
±8 kV air
IEC 61000-4-2
±2 kV for power
Electrical fast
supply lines
transient/burst
±1 kV for input/out-
IEC 61000-4-4
put lines
±1 kV line(s) to
line(s)
Surge
IEC 61000-4-5
±2 kV line(s) to
earth
<5% UT (>95% dip
in UT) for 0,5 cycle
40% UT
Interruptions and
(60% dip in UT)
voltage varia-
for 5 cycles
tions on power
70% UT
supply input lines
(30% dip in UT)
IEC 61000-4-11
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE UT is the A.C. main voltage prior to application of the test level.
Compliance
Electromagnetic Environ-
Level
mental Guidance
Floors should be wood, con-
crete or ceramic tile. If floors
±6 kV contact
are covered with synthetic
±8 kV air
material, the relative humidity
should be at least 30%.
±2 kV for pow-
ersupply lines
Main power quality should be
that of a typical commercial or
±1 kV for input/
hospital environment.
output lines
±1 kV line(s) to
Main power quality should be
line(s)
that of a typical commercial or
±2 kV line(s) to
hospital environment.
earth
<5% UT
(>95% dip in
UT) for 0,5
cycle
Main power quality should be
that of a typical commercial or
40% UT (60%
hospital environment. If the
dip in UT) for 5
user of the model POC-S198
cycles
requires continued operation
during main power interrup-
70% UT (30%
tion, it is recommended that
dip in UT) for
the model POC-S198 be
25 cycles
powered from an uninterrupt-
ible power supply or a battery.
<5% UT
(>95% dip in
UT) for 5 sec
Power frequency magnetic
fields should be at levels
3 A/m
characteristic of a typical
location in a typical commer-
cial or hospital environment.
17
POC-S198 User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Poc-s198-25f-atePoc-s198-bttePoc-s198-bt0e

Table of Contents