Rs232 Command Table; Rs232 Pin Out - OHAUS MB95 Instruction Manual

Moisture analyzer
Hide thumbs Also See for MB95:
Table of Contents

Advertisement

MB95

10.1 RS232 Command Table

Output Formats
Data output can be initiated in one of two ways:
1. By pressing the Print button;
2. Using the Print Interval feature;
RS232 Commands
All communication is accomplished using standard ASCII format. Only the characters
shown in the following table are acknowledged by the Moisture Analyzer. Invalid
command response "ES" error indicates the Moisture Analyzer has not recognized
the command. Commands sent to the Moisture Analyzer must be terminated with a
Line Feed or carriage return-line line feed (CRLF). Data output by the Moisture
Analyzer is always terminated with a carriage return-line feed (CRLF).
Command

10.2 RS232 Pin out

The following table illustrates the pin-out connections on the RS232 connector.
1
2
<–
–>
3
4&6
5
–>
7
8
<–
9
RS232 COMMAND TABLE
Character
Description
V
Print SR Version
TIM
Print Current Time
DAT
Print Current Date
N/C
Date Out (TXD)
Date Out (RXD)
N/C
Ground
Clear to send (CTS)
Request to send (RTS)
N/C
Attention: The RS232 port (COM 1) is located near the
USB device port.
EN-45