7.2.2 Drying temperature

The drying temperature can be set from 40 ° C to 200 ° C.
Attention: If you are working at temperatures above 200 ° C, we advise you to wait 2 to 3
minutes to open the instrument lid between individual measurements to ensure good
reproducibility and avoid overheating of the instrument.

7.2.3 Switch-off criterion

This menu offers you different switch-off criterion. Switch-off criterion defines when the
instrument should end the drying. Switch-off criterion eliminates checking your watch or
clock and stopping the drying manually. Review the list below and select one criterion that
suits your purpose.
Step Profile
For STEP 1, use the defined temperature
and time.
Repeat the above procedure for step 2 and
Final step.
Press the Drying Temperature button to
access the temperature setting menu.
Enter the desired drying temperature and
press the OK button.
The following settings can be selected for the
switch-off criterion:
— Timed switch-off
— Auto switch-off (weight loss per unit of
time - 3 settings)