The user assumes responsibility for any damage caused by the use of these types of samples. Application Disclaimer Moisture determination applications must be optimized and validated by the user according to local regulations. Application specific data provided by OHAUS is for reference purposes only. OHAUS waives all liability for applications based on this data.
EN-2 MB23 / MB25 2. INSTALLATION Package Contents Moisture Analyzer 50 Sample Pans Draft Shield Glass Fiber Pad In-Use Cover Instruction Manual Pan Support Sample Pan Handler (MB25 only) Power Cable Selecting the Location Operate the Analyzer on a firm, level surface.
MB23 / MB25 EN-3 3. OPERATION Display Stable indicator Percent solids Percent moisture grams 100° Temperature setting or current chamber temperature (Celsius) 10:00 Time setting (minutes : seconds) Controls Button: Name: Quick Functions: Button: Name: Quick Functions: On (short press) / Off (long press) Increase value Set ...
Page 6
EN-4 MB23 / MB25 Temperature / Time Edit Mode (Preparing a Test) Temperature editing: (50° to 160° C in 5° steps) Temp (Short press) Enters Edit mode for the temperature setting, the value will blink. Set (Short or long press) Increases the value of the temperature setting.
MB23 / MB25 EN-5 Performing a Trial Test Prior to actual testing, a test run can be performed with these suggested settings: (1) Temperature = 120 (2) Time = AUTO (3) Sample = 3g of water. Place a glass fiber pad (included with the Analyzer) on the test pan, place on Pan Support. Press Tare to zero the pan weight.
Weight calibration is rarely required. Moisture analyzers use relative weight values to determine the results, so a minor offset from the absolute weight has little effect on accuracy. OHAUS moisture analyzers have rugged, high quality temperature stabilized weighing modules that retain their calibration over long periods of time.
MB23 / MB25 EN-7 (7) After 15 minutes, the unit will beep and blink “160” on the display. (8) Read the thermometer on the Temperature Calibration Kit and press Set to change and match the displayed value to the thermometer reading.
Voltage fluctuations: -15% +10% Power load: MB25: 250 W; MB23:300 W (Maximum during drying process) Power line fuse: 1 piece, 5 x 20 mm, 2.5 A 250 V Protected against dust and water, Pollution degree: 2, Installation category: Class II...
Page 11
Compliance to the following standards is indicated by the corresponding mark on the product. Marking Standard This product conforms to the EMC directive 2004/108/EC and the Low Voltage Directive 2006/95/EC. The complete Declaration of Conformity is available from Ohaus Corporation AS/NZS4251.1 Emission, AS/NZS4252.1 CAN/CSA-C22.2 No. 61010-1-04; UL Std. No. 61010A-1...
Page 12
ISO 9001 Registration In 1994, Ohaus Corporation, USA, was awarded a certificate of registration to ISO 9001 by Bureau Veritus Quality International (BVQI), confirming that the Ohaus quality management system is compliant with the ISO 9001 standard’s requirements. On June 21, 2012,...