8.3.10
CO
Analyzer
2
ANALYSIS GRAPH SCALE
BAUDRATE
CO2 ANALYZER RANGE
CO2 ANALYZER CAL
2000ppm
▪ ANALYSIS GRAPH SCALE (2000ppm by default) determines
the scale of the y-axis "CO2 analyzer's ppm-CO2 readings" on
the Analysis Graph Screen (see section
Screen
CO2 data peaks in optimal resolution on the LCD screen.
▪ The ANALYSIS GRAPH SCALE is independent of the CO2
ANALYZER RANGE described below. Note that, in certain
circumstances, when the displayed CO2 peaks on screen
exceeds the scale of the graph, the CO2 analyzer keeps
measuring and integrating the CO2 readings to obtain the TOC
result, without losing any CO2 data.
9600
▪ BAUDRATE (9600bps by default) specifies the data
communication signaling speed of CO
communication interface.
1000ppm
▪ CO2 ANALYZER RANGE defines the full scale range of the CO
analyzer installed in the BioTector.
▪ The CO2 ANALYZER CALIBRATION menu allows the user to
change CO
span parameters when necessary. If required, please contact
manufacturer or distributor for details of the procedures.
for details). This function allows the system to display
Analyzer range and the CO
2
2.1.4 Analysis
Graph
Analyzer with the RS232
2
2
analyzer zero and
2