8.3.10
CO
Analyzer
2
ANALYSIS GRAPH SCALE 10000ppm
BAUDRATE
CO2 ANALYZER RANGE
CO2 ANALYZER CAL
 ANALYSIS GRAPH SCALE determines the scale of the y-axis
"CO2 analyzer's ppm-CO2 readings" on the Analysis Graph
Screen (see section
This function allows the system to display CO2 data peaks in
optimal resolution on the LCD screen.
 The ANALYSIS GRAPH SCALE is independent of the CO2
ANALYZER RANGE described below. Note that, in certain
circumstances, when the displayed CO2 peaks on screen
exceeds the scale of the graph, the CO2 analyzer keeps
measuring and integrating the CO2 readings to obtain the TOC
result, without losing any CO2 data.
 BAUDRATE (9600bps by default) specifies the data
9600
communication signaling speed of CO
communication interface.
 CO2 ANALYZER RANGE defines the full scale range of the CO
10000ppm
analyzer installed in the BioTector.
 The CO2 ANALYZER CALIBRATION menu allows the user to
change CO
span parameters when necessary. If required, please contact
manufacturer or distributor for details of the procedures.
2.1.4 Analysis Graph Screen
Analyzer with the RS232
2
Analyzer range and the CO
2
for details).
2
analyzer zero and
2
Page 132