Hach BioTector B3500ul User Manual page 36

Online toc analyzer
Table of Contents

Advertisement

MONITORING RANGES:
Standard Range
Accuracy and Repeatability
Limit of Detection
SAMPLE & ENVIRONMENTAL CONDITIONS
Sample Volume:
Sample Inlet Pressure:
Drain Pressure:
Sample Inlet Temperature:
Sample Flow Rate:
Sample Particle Size:
Ambient Temperature:
Humidity:
Ingress Protection:
System Sound:
The manufacturer has a continuous research and development program. Specifications may
therefore be changed without notice. For specification updates, please contact the
manufacturer.
TOC
0-5000µgC/l
2% of reading or 10 µgC/l, whichever is greater
10 µg/l
Up to 12.0ml
Ambient (for applications with high sample pressure, sampling systems are
available)
Ambient (for applications with high drain pressure, optional systems are
available)
2°C – 60°C
(36°F - 140°F)
Minimum 100ml per sample
Up to 100 microns
5°C – 45°C
(41°F - 113°F)
Cooling and heating options are available.
5% - 85%, non-condensing
IP44, standard fan cooled, maximum ambient temperature 45°C (113°F)
IP54, air cooled, maximum ambient temperature 35°C (95°F)
IP54, vortex cooled, maximum ambient temperature 50°C (122°F)
< 60 dBa
Page 36

Advertisement

Table of Contents
loading