Flow Test - Hach BioTector B3500ul User Manual

Online toc analyzer
Table of Contents

Advertisement

8.1.1.2

Flow Test

F L O W
T E S T
1 < * E X H A U S T
2
E X H A U S T
3
S A M P L E
4
S A M P L E
T I M E
M F C
S E T P O I N T
M F C
F L O W
S T A T U S
P R E S S
E S C
This menu enables the user to simulate various Flow Tests through the system. The menu also shows the
current status of the Mass Flow Controller.
1. Exhaust Test. Use this function to simulate the flow through the Exhaust Valve. When the Exhaust Test
is activated, an "*" will be shown, and a small menu will display the following data:
Time:
MFC Setpoint:
MFC Flow:
Status:
2. Exhaust Flow. This is similar to the Exhaust Test menu, but its time has been extended to 999s, allowing
the user to locate any blockage in the system. Pass, Warning and Fail are automatically shown on the
screen, depending on the status of the test.
3. Sample Out Test. This is similar to the Exhaust Test. This function allows the user to test the flow
through the Sample Out port (through the Reactor Valve, MV3).
4. Sample Out Flow. This is similar to the Exhaust Flow. This function allows the user to test the flow
through the Sample Out port (through the Reactor Valve, MV3).
T E S T
F L O W
O U T
T E S T
O U T
F L O W
3 0 s
6 0 . 0 l / h
5 8 . 3 l / h
T E S T I N G
T O
A B O R T
The time for the flow test is 30 seconds. This time shows the time left to the end of
the test.
This is the BioTector mass flow controller flow setting (60 l/h by default) for the flow
test.
This is the actual flow from the mass flow controller. If there is no blockage in the
lines, then the setpoint should match the flow.
At the end of the test, the status below is shown:
TESTING:
Test in progress.
PASS:
The Exhaust Test finished its cycle with a flow above the Pass (Flow
Warning) level, which is 45 l/h by default. See Section
WARNING:
The Exhaust Test finished its cycle with a flow below the Pass level
(less than 54 l/h), but above the Fail level (greater than 40 l/h).
FAIL:
The Exhaust Test finished its cycle with a flow below the Fail level,
which is 40 l/h by default.
0 9 : 1 7 : 2 8
1 2 - 0 9 - 0 2
T H E
T E S T
Pressure/Flow Test Program