To ensure that the correct zero offset is applied and to eliminate any remaining TOC
contamination from the analyzer, a Service Zero cycle should be initiated using the SET SERVICE
ZERO function in the Service menu. When initiated, BioTector automatically runs a total of 10 zero
calibration cycles during the following 500 online measurements and automatically adjusts the
zero offset values to compensate for the removal of the contamination. It will therefore not be
necessary to revisit the BioTector after the startup or to repeat the Zero Calibration cycle
manually. See Section
Zero Calibration, ensure that container with at least 5 liters of water is connected to the analyzer. It
is recommended to use a sealed container with CO
on the vent/breathing tube to prevent the contamination of the water.)
Download BioTector "All Data" in text format into the MMC/SD card using the SEND ALL DATA
function in Data Output menu to record all changes made in the system configuration. See Section
8.1.4
Data Output
for details.
Go to Start Stop menu and start the BioTector. When the BioTector is running online, carefully
observe the first two or three reactions and confirm that the CO
8.1.7
Service
for further details. (If ultra-pure water is used during the
absorber and HEPA air filter installed in line
2
peaks are correct.
2