Emergency Stopping

Figure 12A - Pull down on the
RED CORD once to stop the lift.
The unit will beep once and all
power to the lift will be turned off.

Emergency Lowering

Figure 13A - Pull down and hold
until the person is safely lowered to
the desired position.
C450/ C625 - User Guide (753101)
The lift unit also has an Emergency Shut-off feature that allows the operator
to shut the power to the lift unit completely off in the event of an
Emergency. By pulling down once on the RED emergency lowering cord,
located on the underside of the lift unit, the lift will immediately stop and all
its functions will be disabled. The unit will beep once and all power to the
lift will be turned off. The ON Indicator light and display will turn off, and
the Emergency Shut-off button located inside the lift case will pop out. After
an Emergency, the lift unit must be inspected prior to restoring to use. In
Pull
order to restore power back to the lift unit, the white plastic tab that popped
Down
Once
out when the cord was pulled, can be easily pressed back into the lift case by
use of your finger.
Figure 12B – Return power to the lift
unit by pressing in the plastic tab
In the event that the DOWN button on the hand control does not function,
or in power failure situations, the person may be lowered by pulling down
and holding the RED emergency lowering cord located on the underside of
the lift unit. Continue to pull down until the person is safely lowered to the
desired position. The unit will beep as you continue to pull down on the
cord and will continue beeping until the cord is released after the desired
lowering has been achieved. Please refer to figure 13A.
IMPORTANT: The Emergency Lowering button does not provide a rais-
ing function. The failure of any of the lowering devices should be reported
to Prism Medical or your authorized dealer immediately.
Pull
Down
Caution: In an event of emergency when normal lowering system of
and
the lift malfunctions and the
Hold
the lift must be reset by a qualified lift technician before re-use.
Contact your local authorized Prism Medical dealer for service.
Caution: When using the
limit switch will not engage once the strap spools out the maximum 7ft
length. Continued activation will result in reverse spooling and eventually
cause extensive damage to the unit.
"Emergency lowering"
"Emergency lowering"
 
Rev: 14 MAR 2017
Once the RED Emergency Stop/
Lowering Cord is released the lift
unit will need to be reset in order to
operate again. This can be achieved
buy pressing the plastic tab located
at the end of the RED CORD, back
into the lift unit. Then, simply press
any button on the hand control to
resume power. Please refer to Fig-
ure 12B.
function is used ,
function the lower
Page: 22