Emergency Stop; Electrical Emergency Lowering; Manual Emergency Lowering/Raising - Handicare SystemRoMedic RiseAtlas450M QuickTrolleySystem Manual

Table of Contents

Advertisement

Emergency stop

To activate emergency stop: Pull the red emergency stop strap firmly, fig 1, and then release.
To reset: Push the plastic button that holds the red strap in place upward until it clicks,
see fig 2.
When the emergency stop is activated, a red indication is visible, fig 1, above emergency
stop strap and its white plastic button.
If the emergency stop strap is held down constantly, electrical emergency lowering will be
activated. When the strap is released, the lift will.

Electrical emergency lowering

If the hand control is not functioning, electrical emergency lowering
can be activated by holding the emergency stop strap down, fig 1.
Lower the patient to a safe height and location.
Electrical emergency lowering may only be used if the hand
control is malfunctioning. The electrical emergency lowering device
feeds out the lift strap for as long as the emergency stop strap is
held down. Ensure that the lowering motion is activated. Otherwise,
electrical emergency lowering must be stopped. Always ensure that
emergency lowering is done in an appropriate location.
To reset: Press the plastic button that holds the red strap in place
upward until it clicks, see fig 2.

Manual emergency lowering/raising

Remove the plastic cover on the same side as the emergency stop
strap, fig 3. On the inside of the cover there is an Allen key. Remove
the Allen key and insert it into the hole as indicated, fig 4. Turn it
counter-clockwise to lower and clockwise to raise the slingbar.
Always ensure that emergency lowering is done in an appropriate
location.
NOTE! Lowering is very slow.
Re-setting manual emergency lowering/raising: Contact an authorized
Handicare service technician. Do not use the lift unit.
SystemRoMedic
M A N U A L
TM
1
2
3
4
Allen key
17

Hide quick links:

Advertisement

Table of Contents
loading