Cleaning and Disinfection

The exterior of the lift should only be cleaned, disinfected using isopropyl alcohol. Damp a cloth with isopro-
pyl alcohol and wipe down entire exterior of lift and hook. No other chemicals and/or liquids should be used
to clean and disinfect this lift.
protect the lift from liquids may result in damage to the lift and/
or may cause personal injury.
The lift strap may be disinfected using Virox Accel TB RTU (Ready-to-use) available from:
Virox Technologies Inc.
1-800-387-7578
www.virox.com
Details of Parts expected to come in contact with patients during Normal use
The following parts and accessories of C-625/C-450 that are intended to contact the patient in normal use.
 Sling (s).
In rare instances contact of these parts are also possible (not deemed as applied parts )
 Carry Bar.
 Hand Control.
C450/ C625 - User Guide (753101)
Caution: Take great care to ensure that no liquids get inside
the lift. This lift is not drip proof or water tight. Failure to
Rev: 14 MAR 2017