General Inspection and Maintenance
C) Annually - To be completed by a lift technician
This section to be only completed by a qualified service technician as authorized by Prism
Medical.
Annually, the C450/ C625 system must be inspected. It is recommended that this service history is
documented; see pages 34-36 for Service Record History templates.
Complete the inspection as noted in the "Monthly" section above.
Complete the preventative maintenance procedure as outlined in technical manuals for the
C450/ C625 system.
D) Semi-Annually - To be completed by User
In high frequency transfer areas (more than 1500 lifts per year) or situations where heavier than normal clients
regularly are lifted, PM should also be completed every 6 months. It is recommended that this service history
is documented; see pages 34-36 for Service Record History templates.
 
Should any of the these items fail the inspection do not use the lift. Contact your local authorized
dealer for service.
Check for the following:
Check emergency stopping and emergency lowering functions using the RED emergency cord
Inspect carry bar for damage; verify insert and hooks. Ensure no cuts, dents, or sharp edges exist
that may damage the strap of the unit or sling
Lower the lifting tape all the way out of the unit; verify NO signs of fraying or breaking along its
entire length.
Remove the plug from the carry bar and inspect the integrity of the strap around the pin.
The stitching on the lift lifting tape where it connects to the carry bar shows NO signs of fraying,
or breaking.
Verify all functions on the hand control work correctly (UP, DOWN, TRAVERSING, RETURN
TO CHARGE etc.)
The lift has no unusual sounds when the carry bar is moved UP/DOWN or the lift is moved LEFT/
RIGHT.
The airline tube that connects the hand control to the lift is not kinked, twisted, knotted, leaking,
cut or damaged.
Check that the grommet connectors are tight
Hold UP until the carry bar is at the top. Motor should stop automatically.
Using the hand control, let the strap out all the way to the ground. Motor should stop before strap
winds backwards.
While using either UP or DOWN, angle the strap more than 15 degrees. Motor should stop
automatically.
Engage unit with charger and check that unit is charging properly. Light on the lift should turn
amber.
Check that the lift's LED light turns green when on, dark when off, and amber when charging.
Check for cracks or wear in case
Note: if the audible/visual preventative maintenance alert sounds, complete the PM outlined in Section D prior to reset-
ting the lift counter. Instructions to reset the lift counter are detailed on page 21.
C450/ C625 - User Guide (753101)
Rev: 14 MAR 2017
Page: 27