4. Measuring the Concentration

4.3 Measuring using tests without barcode (manual method selection)

When measuring using cell tests without barcode or
reagent tests without AutoSelector, the method must be
selected manually.
The sequence of operations also applies to user-
defined methods.
After approx.
2 s:
12
The last method set up manually appears
on the display.
– Enter the required method number via the numeric
keypad
1
– Confirm with
.
The measured value appears on the display.