Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH Table of contents Table of contents User instructions ............................3 User guide ............................3 1.1.1 Abbreviations ........................3 1.1.2 Symbols ..........................3 Service ............................3 Warranty terms and conditions ......................3 Transportation and storage ......................
Page 4
5.19 Converting the DSEclinical type 5197/5197 eH for right and left-handed users ......43 5.20 Adjusting and operating DentSim ....................44 Service and Care ............................. 45 Maintenance ..........................45 Care ............................... 45 Troubleshooting ............................46 Tested KaVo accessories ........................48 2/48...
12 months from the date of purchase as confirmed by the salesperson. In case of justified complaints, KaVo will honour its warranty with a free replace‐ ment or repair.
Dispose of and recycle the sales packaging appropriately in accordance with cur‐ rent packaging regulations, employing waste management or recycling compa‐ nies. Comply with the comprehensive return system. KaVo has had its sales packaging licensed for this purpose. Please comply with the regional public waste-disposal system.
Conditions, Art. 28). Outside Germany Note KaVo shall not be held liable for damage arising from transportation. The shipment must be checked on arrival. If the packaging is visibly damaged on delivery, please proceed as follows: 1. The recipient of the package must record the loss or damage on the delivery receipt.
Page 8
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 1 User instructions | 1.4 Transportation and storage Permissible stacking load Temperature range Humidity Air pressure 6/48...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 2 Safety | 2.1 Description of safety instructions 2 Safety 2.1 Description of safety instructions 2.1.1 Warning symbol Warning symbol 2.1.2 Structure DANGER The introduction describes the type and source of the hazard. This section describes potential consequences of non-compliance.
▪ Current accident prevention regulations. The overarching guidelines and/or national laws, national regulations and the rules of technology applicable to the startup and use of the KaVo product for the intended purpose are to be applied and complied with. Compliance with the requirements of IEC 62638 (VDE 0701/0702) concerning product safety upon start-up, changes, and repair as well as in repeat examina‐...
2.3 Safety instructions 2.3.1 General DANGER Explosion hazard. Risk of fatal injury. ▶ Do not use KaVo product in areas subject an explosion hazard. DANGER Electricity Electrical shock ▶ Before opening the unit covers, pull the plug from the mains or disconnect the unit from the voltage source.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 2 Safety | 2.3 Safety instructions CAUTION Malfunctions due to electromagnetic fields. The product meets the applicable requirements regarding electromagnetic fields. Given the complex interactions between equipment and cell phones, the product may be influenced by a cell phone that is in use.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 3 Product description | 3.1 DSEclinical Type 5197 / DSEclinical Type 5197 eH optionally available with TM or S table 3 Product description 3.1 DSEclinical Type 5197 / DSEclinical Type 5197 eH optionally available with TM or S table With dentist element TM ①...
Page 14
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 3 Product description | 3.1 DSEclinical Type 5197 / DSEclinical Type 5197 eH optionally available with TM or S table With dentist element S ① Assistant's element ④ Dentist element S ②...
3 Product description | 3.2 Identification plates 3.2 Identification plates Rating plate for the DSEclinical 5197 100 V and rating plate for the DSEclinical 5197 eH 100 V Rating plate for the DSEclinical 5197 120 V and rating plate for the DSEclinical 5197 eH 120 V...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 3 Product description | 3.3 Technical Specifications Follow the instructions for use! Instructions for disposal, see specific purpose - intended use 3.3 Technical Specifications Space needed for DSEclinical type 5197/5197 eH (single chair) Width approximately 486 mm Depth...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 3 Product description | 3.4 Scope of delivery Mode Intermittent operation for Motor K 200 / Operating time is 2 minute; pause is 8 Motor KL 701 / KL 703 minutes Intermittent operation of KL 702 motor The operating time is 0.5 minutes;...
Page 18
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 3 Product description | 3.4 Scope of delivery DSEclinical type 5197 eH basic unit Additional equipment for base unitDSEclinical type 5197 Name Material number BS adjustment electrical DSEclinical E 1.006.3825 Spray mist extractor Description Material number...
Note The DSEclinical type 5197 dental simulation unit may be installed only by trained technical staff and must be bolted to the floor in accordance with KaVo stand‐ ards. The technicians' and fitting instructions for the assembly, setting up and fitting are to be complied with.
Page 20
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 4 First use | 4.1 Fitting and connecting ① PE ③ N ② L1 Requirement The air to be used for the dental simulation unit must be dry, oil-free, non-conta‐ minated, and free of pollution.
Page 21
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 4 First use | 4.1 Fitting and connecting CAUTION Damage caused by non-axial attachment. No media coming through. ▶ Connect axially handpiece hose and handpiece! ▶ Insert the rubber dampers ③ into the gripping sleeve ② of the handpiece. ▶...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.1 Operation in general 5 Operation 5.1 Operation in general 5.1.1 Switch on the DSEclinical type 5197/5197 eH ① Assistant element ③ Main fuse ② Dentist element TM/S ④...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.1 Operation in general Note Once an instrument has been taken out of the holder, the other holders are disa‐ bled. If a second instrument is removed from a holder, it is only recognised after the first instrument is put back.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.2 Overview of controls ▶ Swing the treatment light ① out of the range of movement. ▶ Ensure that the instruments are securely inserted in the holder and tools have been removed from rotating instruments.
Page 25
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.2 Overview of controls Overview of operating elements with dentist element S ① Operating light ⑥ Three-function or multi-function handpiece ② Satelec Mini LED ⑦ Control panel ③...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.3 Control panel CAUTION Rotating instruments and tools Injury hazard ▶ Tie up long hair. ▶ Remove jewellery. ▶ Follow the accident prevention regulations. ▶ Comply with the maximum permissible rotating speed and maximum applica‐ tion pressure in accordance with the regulations of the device manufacturer.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.4 Selecting the dentist 5.4 Selecting the dentist With instruments stored, two dentist levels can be chosen in order to be able to set instrument parameters individually for different types of treatment. ①...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.6 Use INTRA LUX Motor K 200 and INTRA LUX Motor KL 701 / KL 703 5.5.2 Setting the cooling level ▶ Press the preselected spray button. ▶...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.7 Using the PIEZOlux calculus remover The minimum or maximum speed level depends on the motor used and the on the handpiece or angle piece attached. Note The 2 minutes operating time / 5 minutes pause time mode represents the possi‐...
Page 30
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.7 Using the PIEZOlux calculus remover CAUTION Sharp-edged tips. Risk of injury. ▶ When not in use, always keep the supplied torque wrench attached to the tip! 10 N Note The torque wrench (Mat.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.7 Using the PIEZOlux calculus remover 5.7.1 Establishing the cooling level ▶ Regulate the amount of exiting water with the ring. 5.7.2 Changing the high-pressure lamp CAUTION Danger of injury due to hot components Risk of burn injury.
Injury from swallowing the cannula. ▶ Before each treatment, ensure that the cannula is locked into place and firm‐ ly seated. ▶ Only use original KaVo cannulas. Note Cannulas can be rotated by 360°. On-time is 5 minutes while the resting time is 3 minutes.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.9 Using Satelec mini LED ▶ Simultaneously press the air button ① and water button ④ and continuously increase or decrease the exiting spray by applying more or less pressure on the two buttons.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.11 Use the treatment light ▶ Take the required instrument from the instrument holder. 5.11 Use the treatment light Note On completion of treatment, the lamp must always be moved into a safe parked position.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.13 Move the dentist's unit 5.13 Move the dentist's unit Dentist element TM Dentist element S ① Short spring arm ③ Mounting pole ② Short swivel arm The swinging range of the dentist unit is limited by stops.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.14 Changing the height of the assistant element (optional assembly kit) Note Do not pull the dentist unit by the instrument hose. ▶ To adjust the dentist unit height, release the brake, adjust the height, and re‐ set the brake.
Page 37
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.15 Position the patient simulator (pneumatic adjustment) CAUTION Height adjustment of the patient simulator position is associated with a risk of getting pinched. The swivel mechanism moves along when the height is adjusted. ▶...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.16 Electrically position the patient simulator (optional kit) 5.16 Electrically position the patient simulator (optional kit) Note To electrically adjusted the backrest (height and angle), the adjustment kit must be electrically installed Mat.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.16 Electrically position the patient simulator (optional kit) Cross-switch: "Counter‐ Function clockwise motor rotation" The patient simulator moves upward. The patient simulator moves downward. The patient simulator swings upward.
Page 40
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.16 Electrically position the patient simulator (optional kit) The LEDs of the buttons "AP 0", " AP 1", " AP 2" and "SP" flash for about four seconds.
Page 41
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.16 Electrically position the patient simulator (optional kit) with the dentist unit, and two automatic positions can be retrieved with the multi- functional foot control. Retrieve the patient simulator positions with the dentist unit ▶...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.17 Adjust head rest Retrieve the positions when the instrument is removed ▶ Press the stirrup switch and then the "Preselected spray" or "Blown air" foot switch. The patient simulator moves into the automatic position.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.18 Position the head of the patient simulator Removing head cushion The head rest's cushion is detachable. ▶ Loosen the screw. ▶ Pull cushion slightly upwards and then remove it towards the front. 5.18 Position the head of the patient simulator The head of the dental patient simulator can perform the following movements: ▪...
Page 44
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.18 Position the head of the patient simulator Note The head can be drawn out of the torso by approximately 10 cm in order to over- extend the head of the dental patient simulator for specific treatment positions. The headrest must be correctly positioned to prevent the head from slipping backwards unintentionally.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.19 Converting the DSEclinical type 5197/5197 eH for right and left-handed users 5.19 Converting the DSEclinical type 5197/5197 eH for right and left-handed users Requirement The patient simulator is is in the lowest lying position. ▶...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 5 Operation | 5.20 Adjusting and operating DentSim 5.20 Adjusting and operating DentSim DentSim (made by Image Navigation) ① Adjustment/fixation of spring arm ⑤ Fixation of the motion about the lamp-mounting post ②...
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 6 Service and Care | 6.1 Maintenance 6 Service and Care 6.1 Maintenance Note Repair and servicing work on the electrical part of the unit may be done only by skilled staff or by staff trained in the factory.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 7 Troubleshooting 7 Troubleshooting Malfunction Cause Remedy The three-function handpiece or Water pressure or air pressure are ▶ Check pressures and correct, if multifunctional handpiece is not set incorrectly. required. working properly.
Page 49
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 7 Troubleshooting Malfunction Cause Remedy Leaks on instruments O-rings damaged at multiflex/ ▶ Replace O-rings. motor coupling. Water in return air filter. O-rings damaged at multiflex cou‐ ▶ Replace all O-rings in the multi‐ pling.
Instructions for use DSEclinical Typ 5197 / DSEclinical Typ 5197eH 8 Tested KaVo accessories 8 Tested KaVo accessories The accessories listed below are optional and can be supplied depending on unit model. BS = kit Figure Description Material number KL- 701 motor Mat.
Need help?
Do you have a question about the DSEclinical 5197 and is the answer not in the manual?
Questions and answers