Cleaning - Dräger Alcotest 7000 med Instructions For Use Manual

Table of Contents

Advertisement

Available languages

Available languages

WARNING
Fire and explosion hazard
If the device is exposed to sunlight or high ambient temperatures, the
battery may overheat.
► Protect the device from sunlight and high ambient temperatures.
► If the device has been exposed to sunlight or high ambient
temperatures, leave it to cool down.
Measurements cannot be performed during the charging process.
Do not charge the rechargeable battery at temperatures outside the
permissible temperature range of 0 to 45 °C.
Ensure that the power supply unit is easily accessible during
charging.
● Connect the USB-C cable to the USB-C port of the device and a
USB charger.
 The rechargeable battery will charge.
6.2

Cleaning

Clean the device using an alcohol-based cleaning agent.
To clean the device:
1. Turn off the device.
2. If necessary, disconnect the device from the
power supply.
3. Remove the mouthpiece from the device.
Ensure that no liquid gets into the metal
piece of the mouthpiece receptacle on the top
of the device.
4. Wipe the device down with a disposable cloth
with the cleaning agent. Do not apply the
cleaning agent directly to the device.
Dräger Alcotest
®
7000 med
Maintenance
|
en
57