After Use; Work After Testing; Disassembling The Test Unit; Maintenance - Dräger 2500 Instructions For Use Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 16
4.5

After Use

4.5.1

Work after testing

1. Close the compressed air supply.
2. Vent the test unit Disconnect the compressed air cylinder
or the connection to the fixed compressed-air supply from
the test unit.
3. Disassemble the test setup.
4.5.2

Disassembling the test unit

1. Pull the latch of the table mount backwards.
2. Lift the test unit backwards and pull out forward from the
table mount.
5

Maintenance

5.1

Maintenance intervals

Work to do
1)
Perform an inspection
1) Only by trained experts
5.2

Cleaning and disinfecting

Clean the test unit with a moist leather cloth and then dry.
Disinfect the test head with Incides® N disinfection wipes
(order number: 65 70 001)
Do not immerse the test unit in fluids. Do not allow fluids to
come into contact with the connections.
6

Transport

The test unit may not be used during transportation in a
vehicle.
7

Storage

Protect the test unit against coarse contamination and
exposure to hazardous chemicals.
Store the test unit in a dry place. Avoid direct sunlight and
temperatures above 60 °C.
1) Incides is a registered trademark of Ecolab Deutschland GmbH
1)
.
8

Disposal

The Dräger Testor 3500 must not be disposed of as
municipal waste. This is indicated by the adjacent icon.
You can return this product to Dräger free of charge.
For information please contact the national sales
organisations and Dräger.
Batteries and rechargeable batteries must not be
disposed of as municipal waste. This is indicated by
the adjacent icon. Collect and dispose of batteries and
rechargeable batteries at battery collection centres,
in accordance with applicable regulations.
9

Technical data

Ambient conditions
In operation
Temperature
Air pressure
Rel. humidity
X
1)
Weight
Compressed air supply
Medium pressure gauge
Measuring range
Scale division
Class
Low pressure gauge
Measuring range
Scale division
Class
Timer
1) Unit, including timer.
Maintenance
+10 °C to +40 °C
850 mbar to 1400 mbar
max. 70 %
Testor 2500: 5.5 kg
Testor 3500: 6 kg
4 bar to 10 bar
0 bar to 16 bar
0.5 bar
1.6
-15 / 0 / +25 mbar
0.5 mbar
1.0
LCD display
en
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

3500

Table of Contents