Endress+Hauser Float level gauge LT5 Operating Instructions Manual page 73

Mechanical tank gauge for measuring liquid level
Hide thumbs Also See for Float level gauge LT5:
Table of Contents

Advertisement

Float level gauge LT5
Endress+Hauser
1
Ø140 (5.51)
140
130
120
1.
110
100
90
1.
80
70
60
50
40
0.5
 60
Indicator value when the tank is filled with water. Measurement unit: mm (in)
1
Draft surface at density ρ
2
Water (draft surface when the density is 1.0 g (0.002 lb)/cm
Equation: For the φ400 mm (15.75 in) float of LT5-4/LT5-6 8 300 g (18.30 lb)
Float draft when the tank is empty
1.
W - T
V
-
ρ
2
h
Lf =
+
= 80.11mm
2
S
Float diameter
D
Float mass
W
Waterline cross
S
section
Density of the
ρ
measured liquid
Correction of indicated value with simulated actual liquid
2.
Lb = Lf (water) - Lf (actual liquid)
Float diameter
D
Float mass
W
2
 
2.
1.0
400 mm (15.75 in)
8 300 g (18.30 lb)
1 256.64 cm
2
1 g (0.002 lb)/cm
3
(assuming "empty = water")
= 56.50 mm
400 mm (15.75 in)
8 300 g (18.30 lb)
Commissioning
1
Ø140 (5.51)
2.
1.5
3
)
Float height
h
200 mm (7.87 in)
Float volume
V
19 200 cm
Tape tension
T
1 200 g (2.65 lb)
Float height
h
200 mm (7.87 in)
Float volume
V
19 200 cm
 
60
50
40
30
20
10
0
-10
-20
-30
-40
2.0
A0041241
3
3
73

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Float level gauge LT5 and is the answer not in the manual?

Table of Contents