Verify Operation of the Keysight M9383A PXIe Vector Signal Generator

Verify Operation of the Keysight M9383A PXIe Vector Signal
Generator
Before running a Self Test, make sure that all required software is installed, the
chassis is powered on, and all cabling is correct. See
page 32
Allow the M9383A VSG to warm up for at least 30 minutes before using.
1. Select Start > All Programs > Keysight M9383 > M9383 SFP to open the
2. Upon opening the SFP, the Connect to Instrument dialog is displayed. Use
An M9316A module appears in the above dialog as two modules: M9311B and M9319A.
An M9318A module appears in the above dialog as three modules: M9319A, M9348A,
and M9336A.
If not all modules and their slot locations are visible in the SFP Connect to Instrument
dialog, close the SFP and see
Keysight Connection Expert, you may restart the SFP.
3. Check the status of front panel LED states. See
M9383A Startup Guide
for proper cabling.
M9383A SFP.
Control/Select to select all of the modules that are components of the
M9383A and press Connect.
page
57.
"Communications" on page
"Cable The Modules" on
54. After running
"Status LED States" on
53