Check Signal Generator Functionality

Use the following procedure to confirm that the signal generator powers up and that the internal
check identifies no errors. The internal check evaluates the operation of the signal generator and
returns an error message if it detects a problem.
The signal generator must warm up for at least five minutes and all errors must be
cleared from the error queue before a firmware upgrade can be successfully
completed.
If errors exist in the error queue and a firmware upgrade is attempted, the
firmware upgrade will be unsuccessful; new DSP code cannot be upgraded until
the error queue has been cleared.
When signal generators with Option 1E5 are first connected to ac line power, the
error message ERROR 514, Reference Oven Cold occurs, causing both the
OVEN COLD annunciator and the ERR annunciator to turn on.
After approximately five minutes, the OVEN COLD annunciator automatically
clears, but the ERR annunciator remains on until all errors are cleared from the
error queue.
To check signal generator functionality
1. Turn on power to the signal generator and let it warm up for at least five minutes.
2. Cycle power on the signal generator. The green LED should light and the signal generator
should perform a functionality check.
3. When the display lights, check to see if the ERR annunciator is on.
— If the ERR annunciator is on, review the error messages in the error queue by pressing
Utility > Error Info > View Next Error Message. The first error message in the error queue
appears in the display text area. (Refer to the signal generator error messages list on the
documentation CD for information about each error message.)
After resolving all problems causing errors, press Clear Error Queue(s).
— If the ERR annunciator is off, the signal generator functionality check has passed.
Firmware Upgrade Guide E4400-90549
Check Signal Generator Functionality
5