Setting Up Waveform File Vector Modulation - Keysight M9484C Getting Started And Troubleshooting Manual

Vxg vector signal generator
Table of Contents

Advertisement

Basic Measurements
Making Measurements

Setting Up Waveform File Vector Modulation

In this section, we will load a GSM and a LTE waveform into the VXG to
demonstrate the accuracy of the VXG's vector modulation using error vector
magnitude (EVM) measurement applications available on the X-Series Signal
Analyzer.
The VXG supports all ARB waveforms that are provided on the X-Series
sources. This section will use a few of the ARB files that come with the X-Series
sources
Ensure the equipment and triggers are properly configured. Refer to
"Equipment Setup" on page 9
Signal Analyzer" on page 11
Using the graphical user interface
On the VXG:
1. Select Preset > Preset to set the instrument to a known state.
2. Set Frequency to 20 GHz and Amplitude to 0 dBm.
3. Select the Group 1: Signals block to open.
4. Select the arrow for Signal 1 to open the Signal Setup window.
This screen is only accessible if Option M9484C-8SG (8 virtual signal
generators) is installed. For all other option configurations, continue to the
next step.
M9484C Measurement Guide
.
"Setting Up Triggers on the X-Series
and
.
39