% Sat
mg/L
mg/L
mmHg
Salinity
Setup Page 1: View last
calibrated Calibration
Factor
- View the last calibration factor
and 0% offset value.
Note: You will be able to view both Cal
Factor and 0%Sat offset ONLY if you had
Setup Page 2: Select %
Sat mode
- Select to measure DO in %Sat.
Setup Page 3: Set mg/L
mode
- Select to measure DO in mg/L.
Setup Page 4: Set the
Pressure (450 mmHg-
825 mmHg)
- Adjust and select Pressure
value.
Setup Page 5: Set the
Salinity (0.0 - 45.0 ppt)
- Adjust and Select salinity
value.
clear
BUFFER
Overview of Setup Menus in AB 40
9
% Sat
select
°C
S T AB L E
S T AB L E
mg/L
% Sat
Setup Page 6 & 7:
Select number of
calibration points for
%Sat calibration
- Select 1 point or 2 point
calibration for %Saturation.
Setup Page 8 & 9:
Select Auto Calibration
- Set Auto calibration or
Manual calibration for %Sat
and mg/L.
Setup Page 10 & 11:
Select Temperature Unit
- Select unit of measure for
Temperature either in °C or °F.
Setup Page 12:
Enable Stability Indicator
-Set the stability indicator to be
displayed on the screen. Select
On.
Setup Page 13:
Disable Stability
Indicator
-Set the stability indicator to be
NOT displayed on the screen.
Select OFF.
Setup Page 14: Clear
Standardization
- Select to clear standardization.